The biomechanical assessment of fixation methods in periprosthetic femur fractures

Öz

Objective: The aim of our study was to compare the biomechanical properties of different fixation methods used in periprosthetic femur fractures.
Methods: We created sawbone models with Mallory Type 2 periprosthetic femur fractures. The periprosthetic fractures were fixed with cables, cables and strut graft, or cables and plates. The biomechanical properties of these three different fixation methods were compared with the intact femur, the intact femur with prosthesis and the femur with periprosthetic fracture without fixation.
Results: The periprosthetic fracture without fixation had a significantly lower yield point value than the periprosthetic fractures with fixation (p<0.05). There was no significant difference between the three different fixation methods of the periprosthetic fractures. The intact femur with a prosthesis showed statistically higher values than all three fixation methods of perioprosthetic fractures.
Conclusion: Mallory Type 2 periprosthetic fractures should be fixed. There is no difference among the fixation methods used in the study and none of them provide a stability equivalent to that of an intact femur with prosthesis. 

Anahtar Kelimeler

Ayrıntılar

Birincil Dil

İngilizce

Konular

Sağlık Kurumları Yönetimi

Bölüm

Araştırma Makalesi

Yazarlar

Mehmet Gulsen Bu kişi benim

Vasfi Karatosun Bu kişi benim

Bahadır Uyulgan Bu kişi benim

Yayımlanma Tarihi

14 Eylül 2011

Gönderilme Tarihi

7 Mart 2014

Kabul Tarihi

-

Yayımlandığı Sayı

Yıl 2011 Cilt: 45 Sayı: 4

Kaynak Göster

APA
Gulsen, M., Karatosun, V., & Uyulgan, B. (2011). The biomechanical assessment of fixation methods in periprosthetic femur fractures. Acta Orthopaedica et Traumatologica Turcica, 45(4), 266-269. https://izlik.org/JA25JN57EH
AMA
1.Gulsen M, Karatosun V, Uyulgan B. The biomechanical assessment of fixation methods in periprosthetic femur fractures. Acta Orthopaedica et Traumatologica Turcica. 2011;45(4):266-269. https://izlik.org/JA25JN57EH
Chicago
Gulsen, Mehmet, Vasfi Karatosun, ve Bahadır Uyulgan. 2011. “The biomechanical assessment of fixation methods in periprosthetic femur fractures”. Acta Orthopaedica et Traumatologica Turcica 45 (4): 266-69. https://izlik.org/JA25JN57EH.
EndNote
Gulsen M, Karatosun V, Uyulgan B (01 Eylül 2011) The biomechanical assessment of fixation methods in periprosthetic femur fractures. Acta Orthopaedica et Traumatologica Turcica 45 4 266–269.
IEEE
[1]M. Gulsen, V. Karatosun, ve B. Uyulgan, “The biomechanical assessment of fixation methods in periprosthetic femur fractures”, Acta Orthopaedica et Traumatologica Turcica, c. 45, sy 4, ss. 266–269, Eyl. 2011, [çevrimiçi]. Erişim adresi: https://izlik.org/JA25JN57EH
ISNAD
Gulsen, Mehmet - Karatosun, Vasfi - Uyulgan, Bahadır. “The biomechanical assessment of fixation methods in periprosthetic femur fractures”. Acta Orthopaedica et Traumatologica Turcica 45/4 (01 Eylül 2011): 266-269. https://izlik.org/JA25JN57EH.
JAMA
1.Gulsen M, Karatosun V, Uyulgan B. The biomechanical assessment of fixation methods in periprosthetic femur fractures. Acta Orthopaedica et Traumatologica Turcica. 2011;45:266–269.
MLA
Gulsen, Mehmet, vd. “The biomechanical assessment of fixation methods in periprosthetic femur fractures”. Acta Orthopaedica et Traumatologica Turcica, c. 45, sy 4, Eylül 2011, ss. 266-9, https://izlik.org/JA25JN57EH.
Vancouver
1.Mehmet Gulsen, Vasfi Karatosun, Bahadır Uyulgan. The biomechanical assessment of fixation methods in periprosthetic femur fractures. Acta Orthopaedica et Traumatologica Turcica [Internet]. 01 Eylül 2011;45(4):266-9. Erişim adresi: https://izlik.org/JA25JN57EH