The contribution of locked screw-plate fixation with varying angle configurations to stability of osteoporotic fractures: an experimental study

Abstract

Objectives: This experimental study was designed to find new ways of improving stabilization of fractures in osteoporotic elderly patients through alterations made in the configuration and geometry of locked screw-plate fixation used in the conventional plate technique.
Methods: Four screw configurations with varying angulations were used for plate-bone construction. Forty iron plates of high quality (100x35x3 mm) were divided into four groups and two screw holes, 3 mm in diameter, were drilled on each plate at a distance of 15 mm. In group A, the holes were drilled so that the screws would be vertically sent to the bone interface. In the remaining groups, the holes were drilled for convergent (group B, 15°) and divergent (group C, 15°; group D, 30°) screw orientation. Screw-plate fixation was tested in a modified osteoporotic bone (Osteoporotic Generic Bone, Synbone) on an Instron materials testing system with an axial pullout force of 0.1 mm/sec. Failure loads were read from load-displacement curves and the type of failure was noted.
Results: Screws placed in divergent orientations showed the highest axial pull-out strength (group C, 83.3 N/mm; group D, 80.8 N/mm), followed by convergent placement (72 N/mm) and vertical placement (66.7 N/mm). The type of failure was breakage of the bone sample in divergent configurations, and screw pull-out in convergent and vertical configurations.
Conclusion: Divergent constructs may be a promising alternative to conventional screw placement in treating osteoporotic fractures.

Keywords

Ayrıntılar

Birincil Dil

İngilizce

Konular

Sağlık Kurumları Yönetimi

Bölüm

Araştırma Makalesi

Yazarlar

Halil Bekler Bu kişi benim

Guven Bulut Bu kişi benim

Metin Usta Bu kişi benim

Alper Gokce Bu kişi benim

Fethi Okyar Bu kişi benim

Tahsin Beyzadeoglu Bu kişi benim

Yayımlanma Tarihi

12 Haziran 2008

Gönderilme Tarihi

10 Mayıs 2014

Kabul Tarihi

-

Yayımlandığı Sayı

Yıl 2008 Cilt: 42 Sayı: 2

Kaynak Göster

APA
Bekler, H., Bulut, G., Usta, M., Gokce, A., Okyar, F., & Beyzadeoglu, T. (2008). The contribution of locked screw-plate fixation with varying angle configurations to stability of osteoporotic fractures: an experimental study. Acta Orthopaedica et Traumatologica Turcica, 42(2), 125-129. https://izlik.org/JA36DC64HM
AMA
1.Bekler H, Bulut G, Usta M, Gokce A, Okyar F, Beyzadeoglu T. The contribution of locked screw-plate fixation with varying angle configurations to stability of osteoporotic fractures: an experimental study. Acta Orthopaedica et Traumatologica Turcica. 2008;42(2):125-129. https://izlik.org/JA36DC64HM
Chicago
Bekler, Halil, Guven Bulut, Metin Usta, Alper Gokce, Fethi Okyar, ve Tahsin Beyzadeoglu. 2008. “The contribution of locked screw-plate fixation with varying angle configurations to stability of osteoporotic fractures: an experimental study”. Acta Orthopaedica et Traumatologica Turcica 42 (2): 125-29. https://izlik.org/JA36DC64HM.
EndNote
Bekler H, Bulut G, Usta M, Gokce A, Okyar F, Beyzadeoglu T (01 Haziran 2008) The contribution of locked screw-plate fixation with varying angle configurations to stability of osteoporotic fractures: an experimental study. Acta Orthopaedica et Traumatologica Turcica 42 2 125–129.
IEEE
[1]H. Bekler, G. Bulut, M. Usta, A. Gokce, F. Okyar, ve T. Beyzadeoglu, “The contribution of locked screw-plate fixation with varying angle configurations to stability of osteoporotic fractures: an experimental study”, Acta Orthopaedica et Traumatologica Turcica, c. 42, sy 2, ss. 125–129, Haz. 2008, [çevrimiçi]. Erişim adresi: https://izlik.org/JA36DC64HM
ISNAD
Bekler, Halil - Bulut, Guven - Usta, Metin - Gokce, Alper - Okyar, Fethi - Beyzadeoglu, Tahsin. “The contribution of locked screw-plate fixation with varying angle configurations to stability of osteoporotic fractures: an experimental study”. Acta Orthopaedica et Traumatologica Turcica 42/2 (01 Haziran 2008): 125-129. https://izlik.org/JA36DC64HM.
JAMA
1.Bekler H, Bulut G, Usta M, Gokce A, Okyar F, Beyzadeoglu T. The contribution of locked screw-plate fixation with varying angle configurations to stability of osteoporotic fractures: an experimental study. Acta Orthopaedica et Traumatologica Turcica. 2008;42:125–129.
MLA
Bekler, Halil, vd. “The contribution of locked screw-plate fixation with varying angle configurations to stability of osteoporotic fractures: an experimental study”. Acta Orthopaedica et Traumatologica Turcica, c. 42, sy 2, Haziran 2008, ss. 125-9, https://izlik.org/JA36DC64HM.
Vancouver
1.Halil Bekler, Guven Bulut, Metin Usta, Alper Gokce, Fethi Okyar, Tahsin Beyzadeoglu. The contribution of locked screw-plate fixation with varying angle configurations to stability of osteoporotic fractures: an experimental study. Acta Orthopaedica et Traumatologica Turcica [Internet]. 01 Haziran 2008;42(2):125-9. Erişim adresi: https://izlik.org/JA36DC64HM