Investigation of Effect of Air Gap between Surface and Bolus on Dose Distribution for 6 MV Photon Beam

Cilt: 19 Sayı: 4 29 Aralık 2023
PDF İndir
EN

Investigation of Effect of Air Gap between Surface and Bolus on Dose Distribution for 6 MV Photon Beam

Öz

In radiotherapy, tissue equivalent boluses are frequently used in the treatment of superficially located tumors. The air gap between the patient's skin and the bolus may cause dosimetric uncertainties. This study aims to dosimetrically investigate the effect of the air gap between the surface and the bolus on dose distribution. Computed tomography (CT) images of the phantom were obtained and transferred to the treatment planning system (TPS). In the TPS, a bolus was placed on the phantom surface and then air gaps were created between the bolus and the surface. The effect of the air gaps between the surface and the 5 mm thick bolus on the dose distribution was analyzed with the point doses obtained from the TPS. For the 6 MV X-ray, it was observed that the air gap negatively affected the surface doses calculated by TPS. Accordingly, an inverse correlation was found between air gap and surface dose. It is recommended that bolus use, especially in curved anatomical regions, should be applied before CT scanning as much as possible. When using bolus material in radiotherapy, it is recommended to be careful not to leave an air gap between the surface and the bolus.

Anahtar Kelimeler

Kaynakça

  1. [1]. Burnet, N. G. (2004). Defining the tumour and target volumes for radiotherapy. Cancer Imaging, 4(2), 153-161.
  2. [2]. Mukherji, A. (2018). Basics of planning and management of patients during radiation therapy : a guide for students and practitioners. New York, NY: Springer Berlin Heidelberg.
  3. [3]. Park, J. M., Kim, J.-i., Heon Choi, C., Chie, E. K., Kim, I. H., & Ye, S.-J. (2012). Photon energy-modulated radiotherapy: Monte Carlo simulation and treatment planning study. Medical Physics, 39(3), 1265-1277.
  4. [4]. Fadzil, M. S. A., Noor, N. M., Tamchek, N., Ung, N. M., Abdullah, N., Dolah, M. T., & Bradley, D. A. (2022). A cross-validation study of Ge-doped silica optical fibres and TLD-100 systems for high energy photon dosimetry audit under non-reference conditions. Radiation Physics and Chemistry, 200.
  5. [5]. Dogan, N., & Glasgow, G. P. (2003). Surface and build‐up region dosimetry for obliquely incident intensity modulated radiotherapy 6 MV x rays. Medical Physics, 30(12), 3091-3096.
  6. [6]. Zhang, C., Lewin, W., Cullen, A., Thommen, D., & Hill, R. (2023). Evaluation of 3D-printed bolus for radiotherapy using megavoltage X-ray beams. Radiological Physics and Technology, 16(3), 414-421.
  7. [7]. Wang, X., Wang, X., Xiang, Z., Zeng, Y., Liu, F., Shao, B., . . . Liu, L. (2021). The Clinical Application of 3D-Printed Boluses in Superficial Tumor Radiotherapy. Frontiers in Oncology, 11.
  8. [8]. Wang, K. M., Rickards, A. J., Bingham, T., Tward, J. D., & Price, R. G. (2022). Technical note: Evaluation of a silicone‐based custom bolus for radiation therapy of a superficial pelvic tumor. Journal of Applied Clinical Medical Physics, 23(4).

Ayrıntılar

Birincil Dil

İngilizce

Konular

Klasik Fizik (Diğer)

Bölüm

Araştırma Makalesi

Yayımlanma Tarihi

29 Aralık 2023

Gönderilme Tarihi

16 Kasım 2023

Kabul Tarihi

28 Aralık 2023

Yayımlandığı Sayı

Yıl 2023 Cilt: 19 Sayı: 4

Kaynak Göster

APA
Gül, O. V. (2023). Investigation of Effect of Air Gap between Surface and Bolus on Dose Distribution for 6 MV Photon Beam. Celal Bayar University Journal of Science, 19(4), 389-393. https://doi.org/10.18466/cbayarfbe.1391876
AMA
1.Gül OV. Investigation of Effect of Air Gap between Surface and Bolus on Dose Distribution for 6 MV Photon Beam. Celal Bayar University Journal of Science. 2023;19(4):389-393. doi:10.18466/cbayarfbe.1391876
Chicago
Gül, Osman Vefa. 2023. “Investigation of Effect of Air Gap between Surface and Bolus on Dose Distribution for 6 MV Photon Beam”. Celal Bayar University Journal of Science 19 (4): 389-93. https://doi.org/10.18466/cbayarfbe.1391876.
EndNote
Gül OV (01 Aralık 2023) Investigation of Effect of Air Gap between Surface and Bolus on Dose Distribution for 6 MV Photon Beam. Celal Bayar University Journal of Science 19 4 389–393.
IEEE
[1]O. V. Gül, “Investigation of Effect of Air Gap between Surface and Bolus on Dose Distribution for 6 MV Photon Beam”, Celal Bayar University Journal of Science, c. 19, sy 4, ss. 389–393, Ara. 2023, doi: 10.18466/cbayarfbe.1391876.
ISNAD
Gül, Osman Vefa. “Investigation of Effect of Air Gap between Surface and Bolus on Dose Distribution for 6 MV Photon Beam”. Celal Bayar University Journal of Science 19/4 (01 Aralık 2023): 389-393. https://doi.org/10.18466/cbayarfbe.1391876.
JAMA
1.Gül OV. Investigation of Effect of Air Gap between Surface and Bolus on Dose Distribution for 6 MV Photon Beam. Celal Bayar University Journal of Science. 2023;19:389–393.
MLA
Gül, Osman Vefa. “Investigation of Effect of Air Gap between Surface and Bolus on Dose Distribution for 6 MV Photon Beam”. Celal Bayar University Journal of Science, c. 19, sy 4, Aralık 2023, ss. 389-93, doi:10.18466/cbayarfbe.1391876.
Vancouver
1.Osman Vefa Gül. Investigation of Effect of Air Gap between Surface and Bolus on Dose Distribution for 6 MV Photon Beam. Celal Bayar University Journal of Science. 01 Aralık 2023;19(4):389-93. doi:10.18466/cbayarfbe.1391876

Cited By