Calibration of a Platinum Resistance Thermometer (Pt-100) and Its Measurement Uncertainty Analysis
Öz
A common problem found in the industrial metrology is the reliability of
the results obtained from temperature instruments, especially when they are
used in operational processes which effect to the production quality. In this
case, calibration of thermometer associated with a national standard through a
documented unbroken chain of calibrations is of vital importance. This study
presents the reliability of a Pt-100 Platinum Resistance Thermometer (PRT)
measurement results which can be described according to calibration and
statistical methods. The calibrated PRT’s resistance values are characterized
as around 131.00 Ω by making a comparison with a reference PRT at 80 °C bath
temperature and the total uncertainty of the calibration temperature is
determined as ±0.08656 °C by metrological traceability together with
measurement uncertainty. The total uncertainty value changes depending on
uncertainty values of each equipment used in the calibration system.
Anahtar Kelimeler
Kaynakça
- 1. Imai, H., Expanding needs for metrological traceability and measurement uncertainty, Measurement, 2013, 46, 2942–2945.
- 2. Auty, F, Willams, J, Stubbins, R, Brooks, B, Weaver, A, Cunningham M, Cash M, Begineer’s Guide to Measurement in Electronic and Electrical Engineering, National Pysical Laboratory and The Institution of Engineering and Technology (IET), http://www.npl.co.uk/training (accessed 01.08.2017)
- 3. Morris A.S, Measurement and Instrumentation Principle, Butterworth-Heinemann, Linacre House, Jordan Hill, Oxford OX2 8DP, 2001.
- 4. Duvernoy, J, Dubois, A, Instruments and Observing Methods Report No. 86, Training Materials on Metrology and Calibration, World Meteorological Organization, 2006.
- 5. Walker, W.E, Harremoes, P, Rothmans, J, Van Der Sluijs J.P, Van Asselt, M.B.A, Janssen, P, Krayer Von Krauss, M.P, Defining Uncertainty, A Conceptual Basis for Uncertainty Management in Model-Based Decision Support, Integrated Assessment, 2003.
- 6. The International Bureau of Weights and Measures (BIPM), Evaluation of measurement data-Guide to the expression of uncertaintyin measurement, JCGM 100, 2008.
- 7. Kacker, R, Jones, A, On use of Bayesian statistics to make the guide to the expression of uncertainty in measurement consistent, Metrologia, 2003, 40 (5), 235-248.
- 8. Resistance Temperature Detector (RTD) - Principle of Operation, Materials, Configuration and Benefits, Sponsored by Innovative Sensor Technology, USA Division, 2014; https://www.azom.com/article.aspx?ArticleID=5573 (accessed 01.08.2017).
Ayrıntılar
Birincil Dil
İngilizce
Konular
Mühendislik
Bölüm
Araştırma Makalesi
Yazarlar
Levent Paralı
CELÂL BAYAR ÜNİVERSİTESİ
Türkiye
Faruk Durmaz
CELÂL BAYAR ÜNİVERSİTESİ
Olcay Aydın
Bu kişi benim
CELÂL BAYAR ÜNİVERSİTESİ
Yayımlanma Tarihi
30 Mart 2018
Gönderilme Tarihi
16 Ağustos 2017
Kabul Tarihi
5 Mart 2018
Yayımlandığı Sayı
Yıl 2018 Cilt: 14 Sayı: 1