Araştırma Makalesi

Use of A New Extracorporeal Membran Oxygenator System and First Experiences

Cilt: 47 Sayı: 4 30 Aralık 2020
  • Servet Ergün
PDF İndir
TR EN

Use of A New Extracorporeal Membran Oxygenator System and First Experiences

Öz

Objective: In this study, we aimed to share our first experiences about the extracorporeal membrane oxygenator (ECMO) system that we have just started to use and compare it with the system we used before. Methods: Between January 2019 and December 2019, 27 consecutive patients under the age of 18 who were operated for congenital heart disease and needed ECMO support were retrospectively analyzed. İn this period The Medos Deltastream diagonal pump and Sorin Revolution ECMO systems were used. Results: Successful wean was observed in six (42.9%) patients in Medos ECMO group and two (15.4%) in Sorin ECMO group. Five (35.7%) patients discharged in the Medos ECMO group. In Sorin ECMO group all patients died. The difference between the groups in terms of the number of patients discharged was statistically significant (p = 0.02). Lactate clearance was 14 hours in Medos ECMO and 48 hours in Sorin ECMO (p = 0.02). Conclusion: In this study, in which we shared our first experiences regarding the first use of the Sorin ECMO system in the postcardiac pediatric patient group, we observed slower lactate clearance, lower succesful wean and higher mortaly rates. Therefore, this should be taken into account during the use of this system.

Anahtar Kelimeler

Kaynakça

  1. 1. Undar A, Wang S, Arda Ozyuksel A, Rossano JW. Pediatric devices. In: Gregory SD, Stevens MC, Fraser JF. Mechanical Circulatory and Respiratory Support. London: Academic Press. 2017: 271-97.

Ayrıntılar

Birincil Dil

İngilizce

Konular

Sağlık Kurumları Yönetimi

Bölüm

Araştırma Makalesi

Yazarlar

Servet Ergün Bu kişi benim
Türkiye

Yayımlanma Tarihi

30 Aralık 2020

Gönderilme Tarihi

19 Haziran 2020

Kabul Tarihi

1 Aralık 2020

Yayımlandığı Sayı

Yıl 2020 Cilt: 47 Sayı: 4

Kaynak Göster

APA
Ergün, S. (2020). Use of A New Extracorporeal Membran Oxygenator System and First Experiences. Dicle Medical Journal, 47(4), 755-762. https://doi.org/10.5798/dicletip.850129
AMA
1.Ergün S. Use of A New Extracorporeal Membran Oxygenator System and First Experiences. diclemedj. 2020;47(4):755-762. doi:10.5798/dicletip.850129
Chicago
Ergün, Servet. 2020. “Use of A New Extracorporeal Membran Oxygenator System and First Experiences”. Dicle Medical Journal 47 (4): 755-62. https://doi.org/10.5798/dicletip.850129.
EndNote
Ergün S (01 Aralık 2020) Use of A New Extracorporeal Membran Oxygenator System and First Experiences. Dicle Medical Journal 47 4 755–762.
IEEE
[1]S. Ergün, “Use of A New Extracorporeal Membran Oxygenator System and First Experiences”, diclemedj, c. 47, sy 4, ss. 755–762, Ara. 2020, doi: 10.5798/dicletip.850129.
ISNAD
Ergün, Servet. “Use of A New Extracorporeal Membran Oxygenator System and First Experiences”. Dicle Medical Journal 47/4 (01 Aralık 2020): 755-762. https://doi.org/10.5798/dicletip.850129.
JAMA
1.Ergün S. Use of A New Extracorporeal Membran Oxygenator System and First Experiences. diclemedj. 2020;47:755–762.
MLA
Ergün, Servet. “Use of A New Extracorporeal Membran Oxygenator System and First Experiences”. Dicle Medical Journal, c. 47, sy 4, Aralık 2020, ss. 755-62, doi:10.5798/dicletip.850129.
Vancouver
1.Servet Ergün. Use of A New Extracorporeal Membran Oxygenator System and First Experiences. diclemedj. 01 Aralık 2020;47(4):755-62. doi:10.5798/dicletip.850129

Cited By