Use of A New Extracorporeal Membran Oxygenator System and First Experiences

Yıl 2020, Cilt: 47 Sayı: 4, 755 - 762, 30.12.2020
https://doi.org/10.5798/dicletip.850129

Öz

Objective: In this study, we aimed to share our first experiences about the extracorporeal membrane oxygenator (ECMO) system that we have just started to use and compare it with the system we used before.
Methods: Between January 2019 and December 2019, 27 consecutive patients under the age of 18 who were operated for congenital heart disease and needed ECMO support were retrospectively analyzed. İn this period The Medos Deltastream diagonal pump and Sorin Revolution ECMO systems were used.
Results: Successful wean was observed in six (42.9%) patients in Medos ECMO group and two (15.4%) in Sorin ECMO group. Five (35.7%) patients discharged in the Medos ECMO group. In Sorin ECMO group all patients died. The difference between the groups in terms of the number of patients discharged was statistically significant (p = 0.02). Lactate clearance was 14 hours in Medos ECMO and 48 hours in Sorin ECMO (p = 0.02).
Conclusion: In this study, in which we shared our first experiences regarding the first use of the Sorin ECMO system in the postcardiac pediatric patient group, we observed slower lactate clearance, lower succesful wean and higher mortaly rates. Therefore, this should be taken into account during the use of this system.

Kaynakça

  • 1. Undar A, Wang S, Arda Ozyuksel A, Rossano JW. Pediatric devices. In: Gregory SD, Stevens MC, Fraser JF. Mechanical Circulatory and Respiratory Support. London: Academic Press. 2017: 271-97.

Use of A New Extracorporeal Membran Oxygenator System and First Experiences

Yıl 2020, Cilt: 47 Sayı: 4, 755 - 762, 30.12.2020
https://doi.org/10.5798/dicletip.850129

Öz

Objective: In this study, we aimed to share our first experiences about the extracorporeal membrane oxygenator (ECMO) system that we have just started to use and compare it with the system we used before.
Methods: Between January 2019 and December 2019, 27 consecutive patients under the age of 18 who were operated for congenital heart disease and needed ECMO support were retrospectively analyzed. İn this period The Medos Deltastream diagonal pump and Sorin Revolution ECMO systems were used.
Results: Successful wean was observed in six (42.9%) patients in Medos ECMO group and two (15.4%) in Sorin ECMO group. Five (35.7%) patients discharged in the Medos ECMO group. In Sorin ECMO group all patients died. The difference between the groups in terms of the number of patients discharged was statistically significant (p = 0.02). Lactate clearance was 14 hours in Medos ECMO and 48 hours in Sorin ECMO (p = 0.02).
Conclusion: In this study, in which we shared our first experiences regarding the first use of the Sorin ECMO system in the postcardiac pediatric patient group, we observed slower lactate clearance, lower succesful wean and higher mortaly rates. Therefore, this should be taken into account during the use of this system.

Kaynakça

  • 1. Undar A, Wang S, Arda Ozyuksel A, Rossano JW. Pediatric devices. In: Gregory SD, Stevens MC, Fraser JF. Mechanical Circulatory and Respiratory Support. London: Academic Press. 2017: 271-97.
Toplam 1 adet kaynakça vardır.

Ayrıntılar

Birincil Dil İngilizce
Konular Sağlık Kurumları Yönetimi
Bölüm Original Articles
Yazarlar

Servet Ergün Bu kişi benim

Yayımlanma Tarihi 30 Aralık 2020
Gönderilme Tarihi 19 Haziran 2020
Yayımlandığı Sayı Yıl 2020 Cilt: 47 Sayı: 4

Kaynak Göster

APA Ergün, S. (2020). Use of A New Extracorporeal Membran Oxygenator System and First Experiences. Dicle Tıp Dergisi, 47(4), 755-762. https://doi.org/10.5798/dicletip.850129
AMA Ergün S. Use of A New Extracorporeal Membran Oxygenator System and First Experiences. diclemedj. Aralık 2020;47(4):755-762. doi:10.5798/dicletip.850129
Chicago Ergün, Servet. “Use of A New Extracorporeal Membran Oxygenator System and First Experiences”. Dicle Tıp Dergisi 47, sy. 4 (Aralık 2020): 755-62. https://doi.org/10.5798/dicletip.850129.
EndNote Ergün S (01 Aralık 2020) Use of A New Extracorporeal Membran Oxygenator System and First Experiences. Dicle Tıp Dergisi 47 4 755–762.
IEEE S. Ergün, “Use of A New Extracorporeal Membran Oxygenator System and First Experiences”, diclemedj, c. 47, sy. 4, ss. 755–762, 2020, doi: 10.5798/dicletip.850129.
ISNAD Ergün, Servet. “Use of A New Extracorporeal Membran Oxygenator System and First Experiences”. Dicle Tıp Dergisi 47/4 (Aralık 2020), 755-762. https://doi.org/10.5798/dicletip.850129.
JAMA Ergün S. Use of A New Extracorporeal Membran Oxygenator System and First Experiences. diclemedj. 2020;47:755–762.
MLA Ergün, Servet. “Use of A New Extracorporeal Membran Oxygenator System and First Experiences”. Dicle Tıp Dergisi, c. 47, sy. 4, 2020, ss. 755-62, doi:10.5798/dicletip.850129.
Vancouver Ergün S. Use of A New Extracorporeal Membran Oxygenator System and First Experiences. diclemedj. 2020;47(4):755-62.