Araştırma Makalesi

Effects of Pulsatile Flow on Phosphorylcholine Coated Oxygenator and Arterial Filter

Sayı: 34 31 Mart 2022
PDF İndir
EN TR

Effects of Pulsatile Flow on Phosphorylcholine Coated Oxygenator and Arterial Filter

Öz

Objective: We aimed to compare the effects of pulsatile/nonpulsatile flow on phosphorylcholine coated (PC) oxygenator fibers, arterial filters by using scanning electron microscope (SEM). Methods: Eleven patients were randomly divided into two groups, as; nonpulsatile and pulsatile flow groups (NP and P groups) by using PC oxygenators. The oxygenator fiber samples were examined under SEM to compare the thickness of absorbed blood proteins and amount of blood cells on the surface of oxygenators. Arterial filters were also analysed by SEM regarding the captured blood elements or particles. Results: The mean fiber thickness from the axial images were calculated as 46.9 m and 47.6 m at group P and NP respectively which is statistically insignificant. Evaluation of the blood samples that were extracted from the arterial filter bring out higher amount of fibrin network and blood cells on fibers at group NP. Conclusion: We concluded that there is lesser amount of blood components on the fibers of arterial filter at pulsatile flow. Coating of oxygenators is beneficial in case of surface biocompatibility and pulsatile perfusion develops lower amount of blood elements on arterial filter.

Anahtar Kelimeler

Destekleyen Kurum

Destekleyen kurum yoktur.

Proje Numarası

Yoktur.

Teşekkür

Dergi için editöre teşekkürlerimizi bildiririz.

Kaynakça

  1. Abramov, D., Tamariz, M., Serrick, C. I., Sharp, E., Noel, D., Harwood, S., ... & Goldman, B. S. (2003). The influence of cardiopulmonary bypass flow characteristics on the clinical outcome of 1820 coronary bypass patients. The Canadian journal of cardiology, 19(3), 237-243.
  2. Aĝirbaşli, M. A., Song, J., Lei, F., Wang, S., Kunselman, A. R., Clark, J. B., ... & Ündar, A. (2014). Comparative effects of pulsatile and nonpulsatile flow on plasma fibrinolytic balance in pediatric patients undergoing cardiopulmonary bypass. Artificial organs, 38(1), 28-33.
  3. Alghamdi, A. A., & Latter, D. A. (2006). Pulsatile versus nonpulsatile cardiopulmonary bypass flow: an evidence‐based approach. Journal of cardiac surgery, 21(4), 347-354.
  4. Chiu, I. S., Chu, S. H., & Hung, C. R. (1984). Pulsatile flow during routine cardiopulmonary bypass. The Journal of Cardiovascular Surgery, 25(6), 530-536.
  5. De Leval, M., Hill, J. D., Mielke, H., Bramson, M. L., Smith, C., & Gerbode, F. (1972). Platelet kinetics during extracorporeal circulation. ASAIO Journal, 18(1), 355-357.
  6. De Somer, F., Van Belleghem, Y., Foubert, L., François, K., Dubrulle, F., De Wolf, D., & Van Nooten, G. (1960). In vivo evaluation of a phosphorylcholine coated cardiopulmonary bypass. Strategies for optimisation of paediatric cardiopulmonary bypass, 31(2), 241.
  7. Dunn, J., Kirsh, M. M., Harness, J., Carroll, M., Straker, J., & Sloan, H. (1974). Hemodynamic, metabolic, and hematologic effects of pulsatile cardiopulmonary bypass. The Journal of Thoracic and Cardiovascular Surgery, 68(1), 138-147.
  8. Frering, B., Philip, I., Dehoux, M., Rolland, C., Langlois, J. M., & Desmonts, J. M. (1994). Circulating cytokines in patients undergoing normothermic cardiopulmonary bypass. The Journal of thoracic and cardiovascular surgery, 108(4), 636-641.

Ayrıntılar

Birincil Dil

İngilizce

Konular

Mühendislik

Bölüm

Araştırma Makalesi

Yazarlar

Tuna Güray Bu kişi benim
Türkiye

Ece Ürpermez Bu kişi benim
Türkiye

Sertan Özyalçın Bu kişi benim
Türkiye

Yayımlanma Tarihi

31 Mart 2022

Gönderilme Tarihi

5 Mart 2022

Kabul Tarihi

22 Mart 2022

Yayımlandığı Sayı

Yıl 2022 Sayı: 34

Kaynak Göster

APA
Keskin, G., Ulus, A. T., Güray, T., Ürpermez, E., Özyalçın, S., Haberal, O. E., & Kocakulak, M. (2022). Effects of Pulsatile Flow on Phosphorylcholine Coated Oxygenator and Arterial Filter. Avrupa Bilim ve Teknoloji Dergisi, 34, 793-799. https://doi.org/10.31590/ejosat.1083112
AMA
1.Keskin G, Ulus AT, Güray T, vd. Effects of Pulsatile Flow on Phosphorylcholine Coated Oxygenator and Arterial Filter. EJOSAT. 2022;(34):793-799. doi:10.31590/ejosat.1083112
Chicago
Keskin, Gökhan, A. Tulga Ulus, Tuna Güray, vd. 2022. “Effects of Pulsatile Flow on Phosphorylcholine Coated Oxygenator and Arterial Filter”. Avrupa Bilim ve Teknoloji Dergisi, sy 34: 793-99. https://doi.org/10.31590/ejosat.1083112.
EndNote
Keskin G, Ulus AT, Güray T, Ürpermez E, Özyalçın S, Haberal OE, Kocakulak M (01 Mart 2022) Effects of Pulsatile Flow on Phosphorylcholine Coated Oxygenator and Arterial Filter. Avrupa Bilim ve Teknoloji Dergisi 34 793–799.
IEEE
[1]G. Keskin vd., “Effects of Pulsatile Flow on Phosphorylcholine Coated Oxygenator and Arterial Filter”, EJOSAT, sy 34, ss. 793–799, Mar. 2022, doi: 10.31590/ejosat.1083112.
ISNAD
Keskin, Gökhan - Ulus, A. Tulga - Güray, Tuna - Ürpermez, Ece - Özyalçın, Sertan - Haberal, Orhan Erdem - Kocakulak, Mustafa. “Effects of Pulsatile Flow on Phosphorylcholine Coated Oxygenator and Arterial Filter”. Avrupa Bilim ve Teknoloji Dergisi. 34 (01 Mart 2022): 793-799. https://doi.org/10.31590/ejosat.1083112.
JAMA
1.Keskin G, Ulus AT, Güray T, Ürpermez E, Özyalçın S, Haberal OE, Kocakulak M. Effects of Pulsatile Flow on Phosphorylcholine Coated Oxygenator and Arterial Filter. EJOSAT. 2022;:793–799.
MLA
Keskin, Gökhan, vd. “Effects of Pulsatile Flow on Phosphorylcholine Coated Oxygenator and Arterial Filter”. Avrupa Bilim ve Teknoloji Dergisi, sy 34, Mart 2022, ss. 793-9, doi:10.31590/ejosat.1083112.
Vancouver
1.Gökhan Keskin, A. Tulga Ulus, Tuna Güray, Ece Ürpermez, Sertan Özyalçın, Orhan Erdem Haberal, Mustafa Kocakulak. Effects of Pulsatile Flow on Phosphorylcholine Coated Oxygenator and Arterial Filter. EJOSAT. 01 Mart 2022;(34):793-9. doi:10.31590/ejosat.1083112