In this study, a total power radiometer was established to measure the value of "excess noise source" (ENR) of a semiconductor noise source in a traceable manner. The test of this system which can measure the noise source at a time is performed between the frequency range of 50 MHz to 26.5 GHz, accepted as highly wide, by using standard noise sources with a nominal value of 15 dB ENR. The difference between the actual ENR value and the calculated ENR value of the known noise source was obtained as ≤0.30 by using the developed total power radiometer. Uncertainty of the measured ENR values are calculated in the range of 0.13 dB and 0.34 dB, and the difference between the actual value and the measured value was found in the uncertainty values
Excess noise ratio noise noise source total power radiometer uncertainty
Bu çalışmada, yarı iletken bir gürültü kaynağının “excess noise source” (ENR) değerini izlenebilir şekilde ölçmek için bir toplam güç radyometre kurulmuştur. Oldukça geniş olarak kabul edilen 50 MHz – 26,5 GHz frekans aralığında, gürültü kaynağını tek seferde ölçebilen bu sistemin testi, nominal 15 dB ENR değerine sahip standart gürültü kaynakları kullanılarak yapılmıştır. Geliştirilen toplam güç radyometre kullanılarak, değeri bilinen bir gürültü kaynağının hesaplanan ENR değerleri ile gerçek ENR değerleri arasındaki farklar ≤ç0,30ç dB olarak elde edilmiştir. Ölçülen ENR değerlerinin belirsizlikleri ise 0,13 dB ile 0,34 dB aralığında hesaplanmış olup, geçek değer ve ölçülen değer arasındaki farklar belirsizlik değerleri içerisinde bulunmuştur.
In this study, a total power radiometer was established to measure the value of "excess noise source" (ENR) of a semiconductor noise source in a traceable manner. The test of this system which can measure the noise source at a time is performed between the frequency range of 50 MHz to 26.5 GHz, accepted as highly wide, by using standard noise sources with a nominal value of 15 dB ENR. The difference between the actual ENR value and the calculated ENR value of the known noise source was obtained as ≤ç0.30ç by using the developed total power radiometer. Uncertainty of the measured ENR values are calculated in the range of 0.13 dB and 0.34 dB, and the difference between the actual value and the measured value was found in the uncertainty values.
Belirsizlik excess noise ratio gürültü gürültü kaynağı toplam güç radyometre - Excess noise ratio noise noise source total power radiometer uncertainty
Birincil Dil | Türkçe |
---|---|
Bölüm | Akademik ve/veya teknolojik bilimsel makale |
Yazarlar | |
Yayımlanma Tarihi | 14 Şubat 2013 |
Gönderilme Tarihi | 14 Şubat 2013 |
Yayımlandığı Sayı | Yıl 2013 Cilt: 3 Sayı: 5 |
EMO BİLİMSEL DERGİ
Elektrik, Elektronik, Bilgisayar, Biyomedikal, Kontrol Mühendisliği Bilimsel Hakemli Dergisi
TMMOB ELEKTRİK MÜHENDİSLERİ ODASI
IHLAMUR SOKAK NO:10 KIZILAY/ANKARA
TEL: +90 (312) 425 32 72 (PBX) - FAKS: +90 (312) 417 38 18
bilimseldergi@emo.org.tr