Recommending Tolerance Value for SpO2 Devices with Linear Regression Based on Measuring Tape
Öz
Monitoring blood oxygen levels is vital for tracking various respiratory diseases like bronchitis, pneumonia, COPD, and critical care patients, including those with COVID-19. SpO2 devices, calculating oxygen percentage via finger or earlobe tissue, play a crucial role. Field studies have revealed concerns regarding the accuracy of SpO2 measurements due to high deviations, particularly in patients with low oxygen saturation, prompting the initiation of this study to ensure accurate interpretation of the device's measurements. Using a linear regression algorithm, SpO2 values from different bands were classified for quality. Tolerance values and deviation thresholds for each band were recommended. Additionally, linear regression aimed to save time by making result estimations with less data, facilitating more device monitoring and frequent testing. Results closely matched actual values, suggesting contributions to more frequent application and rapid interpretation. Following the European standard 80601-2-61, measurements were taken from three pulse oximeter brands (Contec MS100 model simulator) in three bands: 70-79%, 80-89%, and 90-100% SpO2. For each measurement, an Arms curve graph was generated using linear regression, and the mean square error (MSE) and Arms values were calculated to evaluate devices. In conclusion, deviation rates increase at low oxygen saturation levels, and recommended % SpO2 deviation values were proposed for each band and device quality.
Anahtar Kelimeler
Etik Beyan
There is no need to obtain ethics committee permission for the article prepared. There is no conflict of interest with any person/institution in the prepared article.
Kaynakça
- T. Ş. Eyüboğlu, O. D. Güleryüz, E. Keleş, and I. Budakoğlu, “What do trainee physicians know about the use of pediatric pulse oximetry on patients?, Gazi Med. J., vol. 29, pp. 119–122, 2018.
- R. C. N. McMorrow, and M. G. Mythen, “Pulse oximetry,” Curr. Opin. Crit. Care, vol. 12, pp. 269–271, 2006.
- T. P. Walters, “Pulse oximetry knowledge and its effects on clinical practice,” Br. J. Nurs., vol. 16, no. 21, pp. 1332–1340, 2007.
- M. Elliott, R. Tate, and K. Page, “Do clinicians know how to use pulse oximetry? A literature review and clinical implications,” Aust. Crit. Care, vol. 19, no. 4, pp. 139–144, Nov. 2006.
- K. Jamieson, M. J. Davey, and G. M. Nixon, “Oximetry knowledge in neonatal units in Victoria, Australia: Can we do better?,” Acta Paediatr., vol. 108, no. 3, pp. 468–472, 2019. [6]
- S. Bagha and L. Shaw, “A real-time analysis of PPG signal for measurement of SpO2 and pulse rate,” Int. J. Comput. Appl., vol. 36, no. 11, pp. 45–50, 2015.
- P. A. Assalim, M. T. P. Titisari, D. Caesarendra, and B. A. Prakoso, “Analysis of the effect of red LED and infrared flip flop frequency on SpO2 measurement accuracy,” Indones. J. Electron. Electromed. Eng. Med. Inform., 2022.
- S. Seifi, A. Khatony, G. Moradi, A. Abdi, and F. Najafi, “Accuracy of pulse oximetry in detection of oxygen saturation in patients admitted to the intensive care unit of heart surgery: Comparison of finger, toe, forehead and earlobe probes,” J. Clin. Monit. Comput., vol. 32, no. 6, pp. 1085–1090, 2018.
Ayrıntılar
Birincil Dil
İngilizce
Konular
Yazılım Mühendisliği (Diğer), Biyomedikal Tanı
Bölüm
Araştırma Makalesi
Yayımlanma Tarihi
18 Şubat 2025
Gönderilme Tarihi
27 Ocak 2024
Kabul Tarihi
25 Nisan 2024
Yayımlandığı Sayı
Yıl 2025 Cilt: 4 Sayı: 1
APA
Üstüner, Ö., & Budak, C. (2025). Recommending Tolerance Value for SpO2 Devices with Linear Regression Based on Measuring Tape. Firat University Journal of Experimental and Computational Engineering, 4(1), 1-11. https://doi.org/10.62520/fujece.1426492
AMA
1.Üstüner Ö, Budak C. Recommending Tolerance Value for SpO2 Devices with Linear Regression Based on Measuring Tape. Firat University Journal of Experimental and Computational Engineering. 2025;4(1):1-11. doi:10.62520/fujece.1426492
Chicago
Üstüner, Özge, ve Cafer Budak. 2025. “Recommending Tolerance Value for SpO2 Devices with Linear Regression Based on Measuring Tape”. Firat University Journal of Experimental and Computational Engineering 4 (1): 1-11. https://doi.org/10.62520/fujece.1426492.
EndNote
Üstüner Ö, Budak C (01 Şubat 2025) Recommending Tolerance Value for SpO2 Devices with Linear Regression Based on Measuring Tape. Firat University Journal of Experimental and Computational Engineering 4 1 1–11.
IEEE
[1]Ö. Üstüner ve C. Budak, “Recommending Tolerance Value for SpO2 Devices with Linear Regression Based on Measuring Tape”, Firat University Journal of Experimental and Computational Engineering, c. 4, sy 1, ss. 1–11, Şub. 2025, doi: 10.62520/fujece.1426492.
ISNAD
Üstüner, Özge - Budak, Cafer. “Recommending Tolerance Value for SpO2 Devices with Linear Regression Based on Measuring Tape”. Firat University Journal of Experimental and Computational Engineering 4/1 (01 Şubat 2025): 1-11. https://doi.org/10.62520/fujece.1426492.
JAMA
1.Üstüner Ö, Budak C. Recommending Tolerance Value for SpO2 Devices with Linear Regression Based on Measuring Tape. Firat University Journal of Experimental and Computational Engineering. 2025;4:1–11.
MLA
Üstüner, Özge, ve Cafer Budak. “Recommending Tolerance Value for SpO2 Devices with Linear Regression Based on Measuring Tape”. Firat University Journal of Experimental and Computational Engineering, c. 4, sy 1, Şubat 2025, ss. 1-11, doi:10.62520/fujece.1426492.
Vancouver
1.Özge Üstüner, Cafer Budak. Recommending Tolerance Value for SpO2 Devices with Linear Regression Based on Measuring Tape. Firat University Journal of Experimental and Computational Engineering. 01 Şubat 2025;4(1):1-11. doi:10.62520/fujece.1426492