Recommending Tolerance Value for SpO2 Devices with Linear Regression Based on Measuring Tape

Cilt: 4 Sayı: 1 18 Şubat 2025
PDF İndir
TR EN

Recommending Tolerance Value for SpO2 Devices with Linear Regression Based on Measuring Tape

Öz

Monitoring blood oxygen levels is vital for tracking various respiratory diseases like bronchitis, pneumonia, COPD, and critical care patients, including those with COVID-19. SpO2 devices, calculating oxygen percentage via finger or earlobe tissue, play a crucial role. Field studies have revealed concerns regarding the accuracy of SpO2 measurements due to high deviations, particularly in patients with low oxygen saturation, prompting the initiation of this study to ensure accurate interpretation of the device's measurements. Using a linear regression algorithm, SpO2 values from different bands were classified for quality. Tolerance values and deviation thresholds for each band were recommended. Additionally, linear regression aimed to save time by making result estimations with less data, facilitating more device monitoring and frequent testing. Results closely matched actual values, suggesting contributions to more frequent application and rapid interpretation. Following the European standard 80601-2-61, measurements were taken from three pulse oximeter brands (Contec MS100 model simulator) in three bands: 70-79%, 80-89%, and 90-100% SpO2. For each measurement, an Arms curve graph was generated using linear regression, and the mean square error (MSE) and Arms values were calculated to evaluate devices. In conclusion, deviation rates increase at low oxygen saturation levels, and recommended % SpO2 deviation values were proposed for each band and device quality.

Anahtar Kelimeler

Etik Beyan

There is no need to obtain ethics committee permission for the article prepared. There is no conflict of interest with any person/institution in the prepared article.

Kaynakça

  1. T. Ş. Eyüboğlu, O. D. Güleryüz, E. Keleş, and I. Budakoğlu, “What do trainee physicians know about the use of pediatric pulse oximetry on patients?, Gazi Med. J., vol. 29, pp. 119–122, 2018.
  2. R. C. N. McMorrow, and M. G. Mythen, “Pulse oximetry,” Curr. Opin. Crit. Care, vol. 12, pp. 269–271, 2006.
  3. T. P. Walters, “Pulse oximetry knowledge and its effects on clinical practice,” Br. J. Nurs., vol. 16, no. 21, pp. 1332–1340, 2007.
  4. M. Elliott, R. Tate, and K. Page, “Do clinicians know how to use pulse oximetry? A literature review and clinical implications,” Aust. Crit. Care, vol. 19, no. 4, pp. 139–144, Nov. 2006.
  5. K. Jamieson, M. J. Davey, and G. M. Nixon, “Oximetry knowledge in neonatal units in Victoria, Australia: Can we do better?,” Acta Paediatr., vol. 108, no. 3, pp. 468–472, 2019. [6]
  6. S. Bagha and L. Shaw, “A real-time analysis of PPG signal for measurement of SpO2 and pulse rate,” Int. J. Comput. Appl., vol. 36, no. 11, pp. 45–50, 2015.
  7. P. A. Assalim, M. T. P. Titisari, D. Caesarendra, and B. A. Prakoso, “Analysis of the effect of red LED and infrared flip flop frequency on SpO2 measurement accuracy,” Indones. J. Electron. Electromed. Eng. Med. Inform., 2022.
  8. S. Seifi, A. Khatony, G. Moradi, A. Abdi, and F. Najafi, “Accuracy of pulse oximetry in detection of oxygen saturation in patients admitted to the intensive care unit of heart surgery: Comparison of finger, toe, forehead and earlobe probes,” J. Clin. Monit. Comput., vol. 32, no. 6, pp. 1085–1090, 2018.

Ayrıntılar

Birincil Dil

İngilizce

Konular

Yazılım Mühendisliği (Diğer), Biyomedikal Tanı

Bölüm

Araştırma Makalesi

Yayımlanma Tarihi

18 Şubat 2025

Gönderilme Tarihi

27 Ocak 2024

Kabul Tarihi

25 Nisan 2024

Yayımlandığı Sayı

Yıl 2025 Cilt: 4 Sayı: 1

Kaynak Göster

APA
Üstüner, Ö., & Budak, C. (2025). Recommending Tolerance Value for SpO2 Devices with Linear Regression Based on Measuring Tape. Firat University Journal of Experimental and Computational Engineering, 4(1), 1-11. https://doi.org/10.62520/fujece.1426492
AMA
1.Üstüner Ö, Budak C. Recommending Tolerance Value for SpO2 Devices with Linear Regression Based on Measuring Tape. Firat University Journal of Experimental and Computational Engineering. 2025;4(1):1-11. doi:10.62520/fujece.1426492
Chicago
Üstüner, Özge, ve Cafer Budak. 2025. “Recommending Tolerance Value for SpO2 Devices with Linear Regression Based on Measuring Tape”. Firat University Journal of Experimental and Computational Engineering 4 (1): 1-11. https://doi.org/10.62520/fujece.1426492.
EndNote
Üstüner Ö, Budak C (01 Şubat 2025) Recommending Tolerance Value for SpO2 Devices with Linear Regression Based on Measuring Tape. Firat University Journal of Experimental and Computational Engineering 4 1 1–11.
IEEE
[1]Ö. Üstüner ve C. Budak, “Recommending Tolerance Value for SpO2 Devices with Linear Regression Based on Measuring Tape”, Firat University Journal of Experimental and Computational Engineering, c. 4, sy 1, ss. 1–11, Şub. 2025, doi: 10.62520/fujece.1426492.
ISNAD
Üstüner, Özge - Budak, Cafer. “Recommending Tolerance Value for SpO2 Devices with Linear Regression Based on Measuring Tape”. Firat University Journal of Experimental and Computational Engineering 4/1 (01 Şubat 2025): 1-11. https://doi.org/10.62520/fujece.1426492.
JAMA
1.Üstüner Ö, Budak C. Recommending Tolerance Value for SpO2 Devices with Linear Regression Based on Measuring Tape. Firat University Journal of Experimental and Computational Engineering. 2025;4:1–11.
MLA
Üstüner, Özge, ve Cafer Budak. “Recommending Tolerance Value for SpO2 Devices with Linear Regression Based on Measuring Tape”. Firat University Journal of Experimental and Computational Engineering, c. 4, sy 1, Şubat 2025, ss. 1-11, doi:10.62520/fujece.1426492.
Vancouver
1.Özge Üstüner, Cafer Budak. Recommending Tolerance Value for SpO2 Devices with Linear Regression Based on Measuring Tape. Firat University Journal of Experimental and Computational Engineering. 01 Şubat 2025;4(1):1-11. doi:10.62520/fujece.1426492