Objective: Accurate preoperative localisation is crucial for the surgical management of primary hyperparathyroidism (pHPT). When first-line imaging modalities such as ultrasonography (USG) and Sestamibi SPECT/CT fail to identify the hyperfunctioning gland, four-dimensional computed tomography (4D-CT) may offer additional diagnostic benefit.
Materials and Methods: This retrospective study reviewed 370 parathyroidectomy cases and included adults with primary hyperparathyroidism who underwent 4D-CT after ultrasonography and technetium-99m sestamibi SPECT/CT failed to localise the parathyroid adenoma. Patients with secondary or tertiary hyperparathyroidism were excluded. Demographic and biochemical data, intraoperative findings, and histopathology results were analysed. A standardised three-phase 4D-CT protocol and intraoperative PTH monitoring were used, and the imaging results were correlated with surgical and pathological outcomes to determine diagnostic accuracy.
Results: 4D-CT accurately localised parathyroid adenomas in 13 of 22 patients (59%), with a sensitivity of 68.4%, specificity of 33.3%, positive predictive value of 86.7%, negative predictive value of 14.3%, and overall diagnostic accuracy of 63.6%. Two patients had mediastinal ectopic adenomas, and the false-positive rate was 13.3%. Focused or unilateral neck exploration was feasible in 59% of patients, reducing surgical invasiveness. No major postoperative complications were observed.
Conclusion: 4D-CT provides significant diagnostic and surgical value in patients with nonlocalized pHPT following inconclusive first-line imaging. Its high positive predictive value enables precise localisation and supports minimally invasive parathyroidectomy, improving operative precision and patient outcomes.
Anahtar Kelimeler
Kaynakça
- Liddy S, Worsley D, Torreggiani W, Feeney J. Preoperative imaging in primary hyperparathyroidism: Literature review and recommendations. Can Assoc Radiol J 2017;68(1):47-55. google scholar
- Eller M, Dave A, Johnson C, Fingeret AL. Accuracy of 4-Dimensional computed tomography for localisation in primary hyperparathyroidism. J Surg Res 2021;257:15-21. google scholar
- Raeymaeckers S, De Brucker Y, Vanderhasselt T, Buls N, De Mey J. Detection of parathyroid adenomas with multiphase 4DCT: towards a true four-dimensional technique. BMC Med Imaging 2021;21(1):64. google scholar
- Kurian KE, Hamdi O, Roecker ZA, Barker SA, Strumpf A, Garneau JC, et al. Surgeon review of 4D CT for the identification of parathyroid adenomas. Head Neck. 2025;47(3):839-46. google scholar
- Noltes ME, Kruijff S, Appelman APA, Jansen L, Zandee WT, Links TP, et al. Head-to-head comparison of [11C] methionine PET, [11C] choline PET, and 4-dimensional CT as second-line scans for the detection of parathyroid adenomas in primary hyperparathyroidism. Eur J Nucl Med Mol Imaging 2024;51(4):1050-9. google scholar
- Bellamkonda N, Highland J, McCrary HC, Slattery L, King B, Teames C et al. Fourdimensional computed tomography for parathyroid adenoma localisation: A preoperative imaging protocol. Ann Otol Rhinol Laryngol 2024;133(4):441-8. google scholar
- Scerrino G, Paladino NC, Graceffa G, Melfa G, Orlando G, Di Vuolo R et al. Primary hyperparathyroidism: 18F-fluorocholine PET/CT us 4D-CT for Parathyroid Identification: Towards a Comprehensive Diagnostic Framework-An Updated Review and Recommendations. J Clin Med 2025;14(15):5468. google scholar
- Petranović Ovčariček P, Giovanella L, Hindie E, Huellner MW, Talbot JN, Verburg FA. An essential practice summary of the new EANM guidelines for parathyroid imaging. Q J Nucl Med Mol Imaging. 2022;66(2):93-103. google scholar
Ayrıntılar
Birincil Dil
İngilizce
Konular
Genel Cerrahi, Endokrinoloji
Bölüm
Araştırma Makalesi
Yazarlar
Sezer Akbulut
*
0000-0003-4362-2240
Türkiye
Tuğba Matlım Özel
0000-0001-6305-2957
Türkiye
Aykut Çelik
0000-0002-0478-7723
Türkiye
Görkem Yıldız
0009-0003-5453-6935
Türkiye
Emrecan Deniz
0009-0000-2351-4467
Türkiye
Hamit Yücel Barut
0000-0002-6004-9194
Türkiye
Serkan Sarı
0000-0003-2918-1776
Türkiye
Yayımlanma Tarihi
9 Şubat 2026
Gönderilme Tarihi
3 Haziran 2025
Kabul Tarihi
12 Aralık 2025
Yayımlandığı Sayı
Yıl 2026 Cilt: 89 Sayı: 1