The reliability of a wearable movement analysis system (G-walk) on gait and jump assessment in healthy adults
Öz
Purpose: Wearable inertial sensor systems are generally used in the assessment of gait and jump performance in clinics. G- walk is one of these devices however, before using this device for clinical interpretation, the reliability of this device must be defined. The aim of this study was to investigate the test–retest reliability of the G-walk wearable movement analysis sensor system for gait and jump assessments in healthy adults.
Methods: Forty-nine healthy volunteers (30 females, 19 males, 23.58±2.65 years of age) participated in the study. The jump and gait parameters of the participants were evaluated using G-walk. The gait parameters were; speed, cadence, stride length, stride duration, stance duration, swing duration, double support, single support and pelvic angles during gait. The maximum height of the following jumps was assessed; counter movement jump, squat jump and countermovement jump with arms thrust. The intraclass correlation coefficient (ICC) method was used for test-retest analysis (p<0.05).
Results: All gait parameter measurements had high or excellent test–retest reliability (ICC:0.728-0.969) with the exception of pelvic angles during gait. The assessment of pelvic angles had moderate test–retest reliability (ICC: 0.463-0.659). All of the jump parameters’ measurements had excellent test–retest reliability (ICC: 0.900-0.986).
Conclusion: In conclusion, this study shows that the G-walk is a reliable device for assessing gait and jump. Alongside being portable, easy to use and affordable at cost, the G-walk was found to be reliable in measuring gait and jump performance in healthy adults. The study was registered on clinicaltrials.gov: NCT04310982
Anahtar Kelimeler
Destekleyen Kurum
Proje Numarası
Kaynakça
- 1. Silsupadol P, Teja K, Lugade V. Reliability and validity of a smartphone-based assessment of gait parameters across walking speed and smartphone locations: body, bag, belt, hand, and pocket. Gait Posture. 2017;58:516-522.
- 2. Eltoukhy M, Oh J, Kuenze C, Signorile J. Improved kinect-based spatiotemporal and kinematic treadmill gait assessment. Gait Posture. 2017;51:77-83.
- 3. De Ridder R, Lebleu J, Willems T, et al. Concurrent validity of a commercial wireless trunk triaxial accelerometer system for gait analysis. J Sport Rehabil. 2019;28:6.
- 4. Springer S, Yogev Seligmann G. Validity of the kinect for gait assessment: A focused review. Sensors. 2016;16:194.
- 5. Middleton A, Fritz SL, Lusardi M. Walking speed: the functional vital sign. J Aging Phys Act. 2015;23:314-322.
- 6. Kraan C, Tan A, Cornish K. The developmental dynamics of gait maturation with a focus on spatiotemporal measures. Gait Posture. 2017;51:208-217.
- 7. Markovic G. Does plyometric training improve vertical jump height? A meta-analytical review. Br J Sports Med. 2007;41:349-355.
- 8. Gallardo-Fuentes F, Gallardo-Fuentes J, Ramírez-Campillo R, et al. Intersession and intrasession reliability and validity of the My Jump app for measuring different jump actions in trained male and female athletes. J Strength Cond Res. 2016;30:2049-2056.
Ayrıntılar
Birincil Dil
İngilizce
Konular
Sağlık Kurumları Yönetimi
Bölüm
Araştırma Makalesi
Yazarlar
Gökhan Yazıcı
*
0000-0002-9270-2290
Türkiye
Melek Volkan
0000-0001-9686-0571
Türkiye
Gamze Çobanoğlu
0000-0003-0136-3607
Türkiye
Buse Küpeli
0000-0002-0439-9333
Türkiye
Cagla Ozkul
0000-0001-9367-9910
Türkiye
Deran Oskay
0000-0002-2217-076X
Türkiye
Yayımlanma Tarihi
31 Ağustos 2020
Gönderilme Tarihi
14 Mayıs 2020
Kabul Tarihi
29 Temmuz 2020
Yayımlandığı Sayı
Yıl 2020 Cilt: 7 Sayı: 2