Objective: This in vitro study aimed to evaluate the effect of implant vertical positioning on the accuracy of digital and conventional impressions for single implants using three-dimensional superimposition analysis. Material and Methods: A master model with a single implant analog (Biohorizons, 5.7 mm) was prepared at three different gingival depths (2 mm, 4 mm, and 6 mm) in the mandibular first molar region. Digital impressions were obtained using two intraoral scanners (Trios 3Shape and Cerec Omnicam), while conventional impressions were taken using polyvinylsiloxane (PVS) and condensation silicone (Type C silicone). In total, 96 test models (n=8 per group) were generated. All models were digitized and compared to the reference STL model using reverse-engineering software (Rapidform) for 3D deviation analysis. Statistical analysis was performed using the Kruskal-Wallis and Mann-Whitney U tests (α=0.05). Results: No statistically significant differences were observed between the digital impression groups (p>0.05). Among conventional techniques, the group using C-silicone showed significantly higher deviations than the PVS group (p<0.05). Increasing gingival depth resulted in significantly greater deviation values in all axes, especially in the y- and z-axes at 6 mm depth (p<0.05). Conclusion: The vertical gingival depth has a significant influence on impression accuracy. Digital techniques maintain higher accuracy across varying depths, whereas conventional techniques, particularly those using C-silicone, exhibit reduced performance at deeper levels.

Anahtar Kelimeler

Kaynakça

  1. Karl M, Rosch S, Graef F, Taylor TD, Heckmann SM. Static implant loading caused by as-cast metal and ceramic-veneered superstructures. J Prosthet Dent. 2005;93(4):324-330.
  2. Del’Acqua MA, Chávez AM, Compagnoni MA, Molo F de A. Accuracy of impression techniques for an implant-supported prosthesis. Int J Oral Maxillofac Implants. 2010;25(4):715–721.
  3. Kim Y, Oh TJ, Misch CE, Wang HL. Occlusal considerations in implant therapy: Clinical guidelines with biomechanical rationale. Clinical Oral Implants research, 2005;(1):26-35.
  4. Papaspyridakos P, Chen CJ, Gallucci G, Doukoudakis A, Weber HP, Chronopoulos V. Accuracy of implant impressions for partially and completely edentulous patients: A systematic review. Int J Oral Maxillofac Implants. 2014;29(4):836–845.
  5. Kim JH, Kim KR, Kim S. Critical appraisal of implant impression accuracies: A systematic review. Journal of Prosthetic Dentistry. 2015;114(2):185-192.
  6. Wee AG. Comparison of impression materials for direct multi-implant impressions. J Prosthet Dent. 2000; 83(3):323-331.
  7. Lee SJ, Betensky RA, Gianneschi GE, Gallucci GO. Accuracy of digital versus conventional implant impressions. Clin Oral Implants Res. 2015; 26(6):715-719.
  8. Chochlidakis KM, Papaspyridakos P, Geminiani A, Chen CJ, Feng IJ, Ercoli C. Digital versus conventional impressions for fixed prosthodontics: A systematic review and meta-analysis. Journal of Prosthetic Dentistry. 2016;116(2):184-190.

Ayrıntılar

Birincil Dil

İngilizce

Konular

Sağlık Hizmetleri ve Sistemleri (Diğer)

Bölüm

Araştırma Makalesi

Yayımlanma Tarihi

12 Ağustos 2026

Gönderilme Tarihi

10 Aralık 2025

Kabul Tarihi

2 Şubat 2026

Yayımlandığı Sayı

Yıl 2026 Cilt: 28 Sayı: 2

Kaynak Göster

APA
Aslıyüce, V., Özcan, N., Yiğit İren, E., & Özdemir Önal, B. (2026). IN VITRO EVALUATION OF ACCURACY OF DIGITAL IMPRESSIONS VERSUS CONVENTİONAL IMPRESSIONS AT VARYING IMPLANT DEPTHS FOR SINGLE CROWNS. The Journal of Kırıkkale University Faculty of Medicine, 28(2), 220-226. https://doi.org/10.24938/kutfd.1836746
AMA
1.Aslıyüce V, Özcan N, Yiğit İren E, Özdemir Önal B. IN VITRO EVALUATION OF ACCURACY OF DIGITAL IMPRESSIONS VERSUS CONVENTİONAL IMPRESSIONS AT VARYING IMPLANT DEPTHS FOR SINGLE CROWNS. Kırıkkale Üni Tıp Derg. 2026;28(2):220-226. doi:10.24938/kutfd.1836746
Chicago
Aslıyüce, Vildan, Nihal Özcan, Elif Yiğit İren, ve Buse Özdemir Önal. 2026. “IN VITRO EVALUATION OF ACCURACY OF DIGITAL IMPRESSIONS VERSUS CONVENTİONAL IMPRESSIONS AT VARYING IMPLANT DEPTHS FOR SINGLE CROWNS”. The Journal of Kırıkkale University Faculty of Medicine 28 (2): 220-26. https://doi.org/10.24938/kutfd.1836746.
EndNote
Aslıyüce V, Özcan N, Yiğit İren E, Özdemir Önal B (01 Ağustos 2026) IN VITRO EVALUATION OF ACCURACY OF DIGITAL IMPRESSIONS VERSUS CONVENTİONAL IMPRESSIONS AT VARYING IMPLANT DEPTHS FOR SINGLE CROWNS. The Journal of Kırıkkale University Faculty of Medicine 28 2 220–226.
IEEE
[1]V. Aslıyüce, N. Özcan, E. Yiğit İren, ve B. Özdemir Önal, “IN VITRO EVALUATION OF ACCURACY OF DIGITAL IMPRESSIONS VERSUS CONVENTİONAL IMPRESSIONS AT VARYING IMPLANT DEPTHS FOR SINGLE CROWNS”, Kırıkkale Üni Tıp Derg, c. 28, sy 2, ss. 220–226, Ağu. 2026, doi: 10.24938/kutfd.1836746.
ISNAD
Aslıyüce, Vildan - Özcan, Nihal - Yiğit İren, Elif - Özdemir Önal, Buse. “IN VITRO EVALUATION OF ACCURACY OF DIGITAL IMPRESSIONS VERSUS CONVENTİONAL IMPRESSIONS AT VARYING IMPLANT DEPTHS FOR SINGLE CROWNS”. The Journal of Kırıkkale University Faculty of Medicine 28/2 (01 Ağustos 2026): 220-226. https://doi.org/10.24938/kutfd.1836746.
JAMA
1.Aslıyüce V, Özcan N, Yiğit İren E, Özdemir Önal B. IN VITRO EVALUATION OF ACCURACY OF DIGITAL IMPRESSIONS VERSUS CONVENTİONAL IMPRESSIONS AT VARYING IMPLANT DEPTHS FOR SINGLE CROWNS. Kırıkkale Üni Tıp Derg. 2026;28:220–226.
MLA
Aslıyüce, Vildan, vd. “IN VITRO EVALUATION OF ACCURACY OF DIGITAL IMPRESSIONS VERSUS CONVENTİONAL IMPRESSIONS AT VARYING IMPLANT DEPTHS FOR SINGLE CROWNS”. The Journal of Kırıkkale University Faculty of Medicine, c. 28, sy 2, Ağustos 2026, ss. 220-6, doi:10.24938/kutfd.1836746.
Vancouver
1.Vildan Aslıyüce, Nihal Özcan, Elif Yiğit İren, Buse Özdemir Önal. IN VITRO EVALUATION OF ACCURACY OF DIGITAL IMPRESSIONS VERSUS CONVENTİONAL IMPRESSIONS AT VARYING IMPLANT DEPTHS FOR SINGLE CROWNS. Kırıkkale Üni Tıp Derg. 01 Ağustos 2026;28(2):220-6. doi:10.24938/kutfd.1836746

Bu Dergi, Kırıkkale Üniversitesi Tıp Fakültesi Yayınıdır.