An Investigation of Four-Point Bending Behavior for Personalized Humerus Bone Plate

Abstract

The reverse engineering method has been beneficial for personalized implant designs in the medical field due to its flexible use feature that allows it to be applied in many areas. The method of anatomical features (MAF) is commonly used for fractures of the bone for the development of plate fixation. To improve the geometrical fitting between plate and bone, high-quality image data is the most important factor for computer aided design (CAD ) modelling. This study aims to use MAF for humerus bone plate design that would alternate the fracture region and decrease the stress-shielding effect depending on bending strength. For this purpose, the personalized plate implant (PPI) was designed according to MAF and fabricated using the multi jet fusion (MJF) technique, finally the four-point bending finite element analysis was applied and tested. The results indicated that the PPI structure has higher bending strength than the flat plate design and a much greater surface area.

Keywords

Kaynakça

  1. Ahmad M., Nanda R., Bajwa AS., Candal-Couto J., Green S., Hui AC. Biomechanical testing of the locking compression plate: when does the distance between bone and implant significantly reduce construct stability?. Injury 2007; 38(3): 358-364.
  2. An Z., Zeng B., He X., Chen Q., Hu S. Plating osteosynthesis of mid-distal humeral shaft fractures: minimally invasive versus conventional open reduction technique. International Orthopaedics 2010; 34: 131-135.
  3. Arvier JF., Barker TM., Yau YY., D'Urso PS., Atkinson RL., McDermant GR. Maxillofacial biomodelling. British Journal of Oral and Maxillofacial Surgery 1994; 32(5): 276-283.
  4. ASTM F382-14, Standard specification and test method for metallic bone plates. ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States 2014.
  5. Chen X. Parametric design of patient-specific fixation plates for distal femur fractures. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine 2018; 232(9): 901-911.
  6. Cousley RR., Bainbridge M., Rossouw PE. The accuracy of maxillary positioning using digital model planning and 3D printed wafers in bimaxillary orthognathic surgery. Journal of Orthodontics 2017;44(4): 256-267.
  7. Huotilainen E., Paloheimo M., Salmi M., Paloheimo KS., Björkstrand R., Tuomi, J., Mäkitie A. Imaging requirements for medical applications of additive manufacturing. Acta Radiologica 2014; 55(1): 78-85.
  8. Jabran A., Peach C., Zou Z., Ren L. Parametric design optimisation of proximal humerus plates based on finite element method. Annals of Biomedical Engineering 2019; 47: 601-614.

Ayrıntılar

Birincil Dil

İngilizce

Konular

Biyomekanik Mühendisliği , Katmanlı Üretim

Bölüm

Araştırma Makalesi

Erken Görünüm Tarihi

15 Ocak 2025

Yayımlanma Tarihi

17 Ocak 2025

Gönderilme Tarihi

29 Nisan 2024

Kabul Tarihi

23 Ağustos 2024

Yayımlandığı Sayı

Yıl 2025 Cilt: 8 Sayı: 1

Kaynak Göster

APA
Erdoğuş, H. B. (2025). An Investigation of Four-Point Bending Behavior for Personalized Humerus Bone Plate. Osmaniye Korkut Ata Üniversitesi Fen Bilimleri Enstitüsü Dergisi, 8(1), 167-181. https://doi.org/10.47495/okufbed.1475223
AMA
1.Erdoğuş HB. An Investigation of Four-Point Bending Behavior for Personalized Humerus Bone Plate. Osmaniye Korkut Ata Üniversitesi Fen Bilimleri Enstitüsü Dergisi. 2025;8(1):167-181. doi:10.47495/okufbed.1475223
Chicago
Erdoğuş, Hakan Burçin. 2025. “An Investigation of Four-Point Bending Behavior for Personalized Humerus Bone Plate”. Osmaniye Korkut Ata Üniversitesi Fen Bilimleri Enstitüsü Dergisi 8 (1): 167-81. https://doi.org/10.47495/okufbed.1475223.
EndNote
Erdoğuş HB (01 Ocak 2025) An Investigation of Four-Point Bending Behavior for Personalized Humerus Bone Plate. Osmaniye Korkut Ata Üniversitesi Fen Bilimleri Enstitüsü Dergisi 8 1 167–181.
IEEE
[1]H. B. Erdoğuş, “An Investigation of Four-Point Bending Behavior for Personalized Humerus Bone Plate”, Osmaniye Korkut Ata Üniversitesi Fen Bilimleri Enstitüsü Dergisi, c. 8, sy 1, ss. 167–181, Oca. 2025, doi: 10.47495/okufbed.1475223.
ISNAD
Erdoğuş, Hakan Burçin. “An Investigation of Four-Point Bending Behavior for Personalized Humerus Bone Plate”. Osmaniye Korkut Ata Üniversitesi Fen Bilimleri Enstitüsü Dergisi 8/1 (01 Ocak 2025): 167-181. https://doi.org/10.47495/okufbed.1475223.
JAMA
1.Erdoğuş HB. An Investigation of Four-Point Bending Behavior for Personalized Humerus Bone Plate. Osmaniye Korkut Ata Üniversitesi Fen Bilimleri Enstitüsü Dergisi. 2025;8:167–181.
MLA
Erdoğuş, Hakan Burçin. “An Investigation of Four-Point Bending Behavior for Personalized Humerus Bone Plate”. Osmaniye Korkut Ata Üniversitesi Fen Bilimleri Enstitüsü Dergisi, c. 8, sy 1, Ocak 2025, ss. 167-81, doi:10.47495/okufbed.1475223.
Vancouver
1.Hakan Burçin Erdoğuş. An Investigation of Four-Point Bending Behavior for Personalized Humerus Bone Plate. Osmaniye Korkut Ata Üniversitesi Fen Bilimleri Enstitüsü Dergisi. 01 Ocak 2025;8(1):167-81. doi:10.47495/okufbed.1475223

23487




196541947019414  

1943319434 19435194361960219721 19784  2123822610 23877

* Uluslararası Hakemli Dergi (International Peer Reviewed Journal)

* Yazar/yazarlardan hiçbir şekilde MAKALE BASIM ÜCRETİ vb. şeyler istenmemektedir (Free submission and publication).

* Yılda Ocak, Mart, Haziran, Eylül ve Aralık'ta olmak üzere 5 sayı yayınlanmaktadır (Published 5 times a year)

* Dergide, Türkçe ve İngilizce makaleler basılmaktadır.

*Dergi açık erişimli bir dergidir.

Creative Commons License

Bu web sitesi Creative Commons Atıf 4.0 Uluslararası Lisansı ile lisanslanmıştır.