Evaluation of the TRK-2P Instrument Reliability in Normal and Keratoconus Eyes: A Preliminary Observation
Öz
To compare keratometry and corneal thickness measurements by the TRK-2P instrument (Topcon Medical Systems Inc., Oakland NJ) with the anterior cornea keratometry and pachymetry values obtained by the Pentacam-HR instrument (Oculus; Optikgeräte GmbH, Wetzlar, Germany). Patients who had full records of two both instruments in our databases were included in the observational study. Keratoconus diagnosed twenty-three eyes of twelve patients and thirty-two eyes of sixteen patients with no eye problem (controls) were included. The keratometry and the central corneal thickness (CCT) outputs were collected by TRK-2P and Pentacam-HR. The consistency of the mean anterior cornea keratometry and pachymetry data were correlated using the intraclass correlation coefficient (ICC). Means were statistically correlated using a Paired t-test seeking significant correlations (α=0.05). Mean keratometry of TRK-2P and Pentacam-HR were 42.64D±2.02 and 42.79D±1.95 in controls whereas, these were 47.64D±5.24 and 47.16D±4.65 in keratoconus, respectively. The mean differences in keratometry data were 0.14D for controls and 0.48D for keratoconus(p<0.001). Mean CCT of TRK-2P and Pentacam-HR were 560.27±42.18µm and 537.63±36µm in controls whereas, these were 489.67±45.13µm and 470.22±38.14µm in keratoconus, respectively. The mean differences in CCT data were 22.63µm for controls and 19.44µm for keratoconus(p<0.001). ICC values between two instruments for controls and keratoconus, respectively were as follows: 0.987 and 0.983 for keratometry, 0.998 and 0.994 for CCT (p<0.001). TRK-2P produces consistent result outputs in normal and pathological corneas. Also, TRK-2P is a reliable instrument when correlated with a reference high-reliable instrument. However, in terms of monitoring the progression of keratoconus, these instruments cannot be interchangeable alternatives
Anahtar Kelimeler
Kaynakça
- 1. Rabinowitz YS. Keratoconus. Surv Ophthalmol. 1998;42:297-319.
- 2. Li Y, Meisler DM, Tang M, et al. Keratoconus diagnosis with optical coherence tomography pachymetry mapping. Ophthalmology. 2008;115:2159-66.
- 3. Ciolino JB, Khachikian SS, Belin MW. Comparison of corneal thickness measurements by ultrasound and scheimpflug photography in eyes that have undergone laser in situ keratomileusis. Am J Ophthalmol. 2008;145:75-80.
- 4. McAlinden C, Khadka J, Pesudovs K. A comprehensive evaluation of the precision (repeatability and reproducibility) of the Oculus Pentacam HR. Invest Ophthalmol Vis Sci. 2011 29;52:7731-7.
- 6. Topcon Medical Systems. [Internet]. TRK-2P Product brochure; [Cited 2021 April 20] Available from: https://www.topcon-medical.eu/files/EU_Downloads/Products/TRK-2P/TRK-2P_brochure_en.pdf
- 5. Reinstein DZ, Gobbe M, Archer TJ. Anterior segment biometry: a study and review of resolution and repeatability data. J Refract Surg. 2012;28:509-20.
- 7. Ozyol E, Özyol P. Comparison of central corneal thickness with four noncontact devices: An agreement analysis of swept-source technology. Indian J Ophthalmol. 2017;65:461-5.
- 8. Kocamis O, Kilic R. Repeatability, reproducibility and agreement of central corneal thickness measurements by two noncontact pachymetry devices. Med Hypothesis Discov Innov Ophthalmol. 2019;8:34-39.
Ayrıntılar
Birincil Dil
İngilizce
Konular
Sağlık Kurumları Yönetimi
Bölüm
Araştırma Makalesi
Yayımlanma Tarihi
28 Kasım 2022
Gönderilme Tarihi
29 Nisan 2022
Kabul Tarihi
7 Haziran 2022
Yayımlandığı Sayı
Yıl 2022 Cilt: 44 Sayı: 6
APA
Tuncer Orhan, F., Atalay, E., & Yıldırım, N. (2022). Evaluation of the TRK-2P Instrument Reliability in Normal and Keratoconus Eyes: A Preliminary Observation. Osmangazi Tıp Dergisi, 44(6), 773-778. https://doi.org/10.20515/otd.1111360
AMA
1.Tuncer Orhan F, Atalay E, Yıldırım N. Evaluation of the TRK-2P Instrument Reliability in Normal and Keratoconus Eyes: A Preliminary Observation. Osmangazi Tıp Dergisi. 2022;44(6):773-778. doi:10.20515/otd.1111360
Chicago
Tuncer Orhan, Feride, Eray Atalay, ve Nilgün Yıldırım. 2022. “Evaluation of the TRK-2P Instrument Reliability in Normal and Keratoconus Eyes: A Preliminary Observation”. Osmangazi Tıp Dergisi 44 (6): 773-78. https://doi.org/10.20515/otd.1111360.
EndNote
Tuncer Orhan F, Atalay E, Yıldırım N (01 Kasım 2022) Evaluation of the TRK-2P Instrument Reliability in Normal and Keratoconus Eyes: A Preliminary Observation. Osmangazi Tıp Dergisi 44 6 773–778.
IEEE
[1]F. Tuncer Orhan, E. Atalay, ve N. Yıldırım, “Evaluation of the TRK-2P Instrument Reliability in Normal and Keratoconus Eyes: A Preliminary Observation”, Osmangazi Tıp Dergisi, c. 44, sy 6, ss. 773–778, Kas. 2022, doi: 10.20515/otd.1111360.
ISNAD
Tuncer Orhan, Feride - Atalay, Eray - Yıldırım, Nilgün. “Evaluation of the TRK-2P Instrument Reliability in Normal and Keratoconus Eyes: A Preliminary Observation”. Osmangazi Tıp Dergisi 44/6 (01 Kasım 2022): 773-778. https://doi.org/10.20515/otd.1111360.
JAMA
1.Tuncer Orhan F, Atalay E, Yıldırım N. Evaluation of the TRK-2P Instrument Reliability in Normal and Keratoconus Eyes: A Preliminary Observation. Osmangazi Tıp Dergisi. 2022;44:773–778.
MLA
Tuncer Orhan, Feride, vd. “Evaluation of the TRK-2P Instrument Reliability in Normal and Keratoconus Eyes: A Preliminary Observation”. Osmangazi Tıp Dergisi, c. 44, sy 6, Kasım 2022, ss. 773-8, doi:10.20515/otd.1111360.
Vancouver
1.Feride Tuncer Orhan, Eray Atalay, Nilgün Yıldırım. Evaluation of the TRK-2P Instrument Reliability in Normal and Keratoconus Eyes: A Preliminary Observation. Osmangazi Tıp Dergisi. 01 Kasım 2022;44(6):773-8. doi:10.20515/otd.1111360