Transcatheter Closure of Patent Ductus Arteriosus in infants between 2-10 kg
Öz
Objective: The aim of this study was to present our experiences on transcatheter patent ductus arteriosus (PDA) closure with different devices, mostly the Amplatzer Piccolo Occluder, in infants weighing between 2-10 kg.
Material and Methods: In this study, the files of 31 patients who underwent transcatheter PDA closure between December 2019 and August 2022 were reviewed retrospectively.
Results: Transcatheter PDA closure was performed on 31(14 female) infants weighing between 2-10 kg. The mean age of patients was 10.7±6.7 months (2-28), and the mean weight was 6.6±2 kg (3-9.9). The median narrowest diameter of the ductus was 2.2 mm (IQR 2-3) and the median ductus length was 6 mm (IQR 4.75-8). The procedural success rate of all interventional procedures was 88% (30 of 34). Complications occurred in a total of four patients including failure of device implantation in one patient, post-procedural device embolization in 2 patients, and the significant residual shunt in one patient. None of the patients required surgery. In 34 interventional procedures, 3 of which were reintervention, 34 devices were used. Twenty-seven (79%) of them were Amplatzer Piccolo Occluder. The median fluoroscopy and procedural times were 10.5 minutes (IQR 7.25-18.5) and 40 minutes (IQR 35-57.5) respectively. The mean duration of follow-up was 10.3±8.8 months (1-32 months).
Conclusion: In our experience, transcatheter treatment of PDA with the Amplatzer Piccolo Occluder device which was our first choice for appropriate duct anatomy and size in infants weighing between 2-10 kg, is safe and effective.
Anahtar Kelimeler
Kaynakça
- Latson LA, Hofschire PJ, Kugler JD, Cheatham JP, Gumbiner CH, Danford DA. Transcatheter closure of patent ductus arteriosus in pediatric patients. J Pediatr 1989;115:549-53.
- Sathanandam SK, Gutfinger D, O’Brien L, Forbes T J, Gillespie MJ, Berman DP, et al. Amplatzer Piccolo Occluder clinical trial for percutaneous closure of the patent ductus arteriosus in patients ≥700 grams. Catheter Cardiovasc Interv 2020;96:1266-76
- Bhat YA, Almesned A, Alqwaee A, Al Akhfash A. Catheter Closure of Clinically Silent Patent Ductus Arteriosus Using the Amplatzer Duct Occluder II-Additional Size: A Single-Center Experience. Cureus 2021;13:e17481.
- Santoro G, Giordano M, Gaio G, Palladino MT, Capozzi G, Iacono C et al. Transcatheter Closure of Arterial Duct in Infants < 6 kg: Amplatzer Duct Occluder Type I vs Amplatzer Duct Occluder II Additional Sizes. Pediatr Cardiol 2018;39:627-32.
- Barcroft M, McKee C, Berman DP, Taylor RA, Rivera BK, Smith CV et al. Percutaneous Closure of Patent Ductus Arteriosus. Clin Perinatol 2022;49:149-66.
- Morville P, Akhavi A. Transcatheter closure of hemodynamic significant patent ductus arteriosus in 32 premature infants by amplatzer ductal occluder additional size-ADOIIAS. Catheter Cardiovasc Interv 2017;90:612-7.
- Pamukcu O, Tuncay A, Narin N, Baykan A, Korkmaz L, Argun M et al. Patent Ductus Arteriosus closure in preterms less than 2kg: Surgery versus transcatheter. Int J Cardiol 2018;250:110-15.
- Baspinar O, Sahin DA, Sulu A, Irdem A, Gokaslan G, Sivasli E, et al. Transcatheter closure of patent ductus arteriosus in under 6 kg and premature infants. J Interv Cardiol 2015;28:180-9.
Ayrıntılar
Birincil Dil
İngilizce
Konular
İç Hastalıkları
Bölüm
Araştırma Makalesi
Yazarlar
Seçil Sayın
0000-0001-6581-6121
Türkiye
Bilal Özelce
0000-0002-9107-4894
Türkiye
Denizhan Bağrul
0000-0003-0375-1726
Türkiye
Alper Gürsu
0000-0002-0707-2678
Türkiye
İbrahim Ece
0000-0002-3657-2209
Türkiye
Yayımlanma Tarihi
30 Ocak 2023
Gönderilme Tarihi
17 Ekim 2022
Kabul Tarihi
28 Kasım 2022
Yayımlandığı Sayı
Yıl 2023 Cilt: 17 Sayı: 1