Preliminary Evaluation of Interocular Agreement and Biometric Associations of Cone Photoreceptor Metrics in Healthy Young Adults Using Confocal Retinal Imaging

Cilt: 52 6 Ekim 2026
PDF İndir
TR EN

Preliminary Evaluation of Interocular Agreement and Biometric Associations of Cone Photoreceptor Metrics in Healthy Young Adults Using Confocal Retinal Imaging

Öz

This study evaluated the interocular agreement and within-eye repeatability of cone photoreceptor metrics measured using a confocal cone analysis system in healthy young adults, as well as their associations with ocular dominance and axial length. This cross-sectional study included 94 eyes from 47 healthy participants. Cone density and cone diameter were measured from fovea-centered 3 × 3 mm images obtained using the MicroClear CRO system (MicroClear Biomedical Co., Ltd., Suzhou, China). Magnification-corrected and original uncorrected cone measurements were analyzed. Interocular agreement was assessed using paired comparisons, intraclass correlation coefficients (ICC), and Bland–Altman analysis. Within-eye repeatability was evaluated from two repeated measurements per eye using the within-subject standard deviation, coefficient of repeatability (CoR), ICC, and Bland–Altman analysis. Associations with axial length were examined using Spearman correlation and generalized estimating equation (GEE) models. No significant interocular differences were observed for magnification-corrected cone density (mean difference, −763 cells/mm²; 95% confidence interval [CI], −2,799 to 1,273; p = 0.455; ICC = 0.679) or cone diameter (mean difference, 0.07 μm; 95% CI, −0.12 to 0.27; p = 0.455; ICC = 0.359). After exclusion of four measurements with automated segmentation failure, within-eye repeatability was good for both cone density (CoR = 4,937 cells/mm²; ICC = 0.924) and cone diameter (CoR = 0.40 μm; ICC = 0.941). Cone photoreceptor parameters were also similar between dominant and nondominant eyes. At the participant level, magnification-corrected cone density showed a moderate negative correlation with axial length (Spearman ρ = −0.545; p < 0.001), and this association remained significant in the adjusted GEE model (β = −5,914.39 cells/mm²/mm; 95% CI, −8,567.73 to −3,261.06; p < 0.001). In contrast, axial length was not significantly associated with uncorrected cone density in either the participant-level correlation analysis (Spearman ρ = −0.121; p = 0.419) or the adjusted GEE model (β = −1,798.92 cells/mm²/mm; 95% CI, −4,104.46 to 506.62; p = 0.126). These findings indicate good within-eye repeatability after exclusion of segmentation-failure measurements, while the association between axial length and cone density was sensitive to the magnification-correction procedure and should therefore be interpreted cautiously.

Anahtar Kelimeler

Etik Beyan

Ethics Committee Decision No. 97, Decision Date: August 27, 2025

Kaynakça

  1. 1. Behar-Cohen F, Gelize E, Jonet L, et al. [Anatomy of the retina]. Med Sci (Paris) 2020; 36: 594–599. 20200702. DOI: 10.1051/medsci/2020094.
  2. 2. Lamb TD. Why rods and cones? Eye (Lond) 2016; 30: 179–185. 20151113. DOI: 10.1038/eye.2015.236.
  3. 3. Zhang T, Godara P, Blanco ER, et al. Variability in Human Cone Topography Assessed by Adaptive Optics Scanning Laser Ophthalmoscopy. Am J Ophthalmol 2015; 160: 290–300 e291. 20150430. DOI: 10.1016/j.ajo.2015.04.034.
  4. 4. Zhou WD, Dong L, Shi HX, et al. Cone mosaic in eyes with varied axial length using adaptive optics scanning laser ophthalmoscopy. Photodiagnosis Photodyn Ther 2025; 52: 104517. 20250208. DOI: 10.1016/j.pdpdt.2025.104517.
  5. 5. Dabir S, Mangalesh S, Schouten JS, et al. Axial length and cone density as assessed with adaptive optics in myopia. Indian J Ophthalmol 2015; 63: 423–426. DOI: 10.4103/0301-4738.159876.
  6. 6. Tumahai P, Moureaux C, Meillat M, et al. High-resolution imaging of photoreceptors in healthy human eyes using an adaptive optics retinal camera. Eye (Lond) 2018; 32: 1723–1730. 20180711. DOI: 10.1038/s41433-018-0140-1.
  7. 7. Liang J, Fang D, Diao Y, et al. Photoreceptor cone loss is associated with decreased deep retinal vessel density and impaired retinal sensitivity in high myopia. Br J Ophthalmol 2025; 109: 1194–1200. 20250923. DOI: 10.1136/bjo-2024-326609.
  8. 8. Wang Y, Bensaid N, Tiruveedhula P, et al. Human foveal cone photoreceptor topography and its dependence on eye length. Elife 2019; 8 20190726. DOI: 10.7554/eLife.47148.

Ayrıntılar

Birincil Dil

İngilizce

Konular

Göz Hastalıkları

Bölüm

Araştırma Makalesi

Yayımlanma Tarihi

6 Ekim 2026

Gönderilme Tarihi

8 Haziran 2026

Kabul Tarihi

12 Eylül 2026

Yayımlandığı Sayı

Yıl 2026 Cilt: 52

Kaynak Göster

APA
Karaca, S., Kızılay, O., Kırık, F., Batu Oto, B., & Çelik, G. (2026). Preliminary Evaluation of Interocular Agreement and Biometric Associations of Cone Photoreceptor Metrics in Healthy Young Adults Using Confocal Retinal Imaging. Journal of Uludağ University Medical Faculty, 52. https://doi.org/10.32708/uutfd.1966222
AMA
1.Karaca S, Kızılay O, Kırık F, Batu Oto B, Çelik G. Preliminary Evaluation of Interocular Agreement and Biometric Associations of Cone Photoreceptor Metrics in Healthy Young Adults Using Confocal Retinal Imaging. Uludağ Üniv Tıp Fak Derg. 2026;52. doi:10.32708/uutfd.1966222
Chicago
Karaca, Serap, Osman Kızılay, Furkan Kırık, Bilge Batu Oto, ve Gökhan Çelik. 2026. “Preliminary Evaluation of Interocular Agreement and Biometric Associations of Cone Photoreceptor Metrics in Healthy Young Adults Using Confocal Retinal Imaging”. Journal of Uludağ University Medical Faculty 52 (Ekim). https://doi.org/10.32708/uutfd.1966222.
EndNote
Karaca S, Kızılay O, Kırık F, Batu Oto B, Çelik G (01 Ekim 2026) Preliminary Evaluation of Interocular Agreement and Biometric Associations of Cone Photoreceptor Metrics in Healthy Young Adults Using Confocal Retinal Imaging. Journal of Uludağ University Medical Faculty 52
IEEE
[1]S. Karaca, O. Kızılay, F. Kırık, B. Batu Oto, ve G. Çelik, “Preliminary Evaluation of Interocular Agreement and Biometric Associations of Cone Photoreceptor Metrics in Healthy Young Adults Using Confocal Retinal Imaging”, Uludağ Üniv Tıp Fak Derg, c. 52, Eki. 2026, doi: 10.32708/uutfd.1966222.
ISNAD
Karaca, Serap - Kızılay, Osman - Kırık, Furkan - Batu Oto, Bilge - Çelik, Gökhan. “Preliminary Evaluation of Interocular Agreement and Biometric Associations of Cone Photoreceptor Metrics in Healthy Young Adults Using Confocal Retinal Imaging”. Journal of Uludağ University Medical Faculty 52 (01 Ekim 2026). https://doi.org/10.32708/uutfd.1966222.
JAMA
1.Karaca S, Kızılay O, Kırık F, Batu Oto B, Çelik G. Preliminary Evaluation of Interocular Agreement and Biometric Associations of Cone Photoreceptor Metrics in Healthy Young Adults Using Confocal Retinal Imaging. Uludağ Üniv Tıp Fak Derg. 2026;52. doi:10.32708/uutfd.1966222.
MLA
Karaca, Serap, vd. “Preliminary Evaluation of Interocular Agreement and Biometric Associations of Cone Photoreceptor Metrics in Healthy Young Adults Using Confocal Retinal Imaging”. Journal of Uludağ University Medical Faculty, c. 52, Ekim 2026, doi:10.32708/uutfd.1966222.
Vancouver
1.Serap Karaca, Osman Kızılay, Furkan Kırık, Bilge Batu Oto, Gökhan Çelik. Preliminary Evaluation of Interocular Agreement and Biometric Associations of Cone Photoreceptor Metrics in Healthy Young Adults Using Confocal Retinal Imaging. Uludağ Üniv Tıp Fak Derg. 01 Ekim 2026;52. doi:10.32708/uutfd.1966222

ISSN: 1300-414X, e-ISSN: 2645-9027

Uludağ Üniversitesi Tıp Fakültesi Dergisi "Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License" ile lisanslanmaktadır.


Creative Commons License
Journal of Uludag University Medical Faculty is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.

2023