Icare Tonometresi ve Goldmann Aplanasyon Tonometrelerinin Klinik Olarak Karşılaştırılması

Cilt: 37 Sayı: 2 1 Haziran 2011
  • Mehmet Baykara
  • Sevil Türüdü
  • Nükhet Zaım
  • Allahyar Ağayarov
  • Hikmet Özçetin
PDF İndir
EN TR

Clinical Comparison of Icare Tonometer and Goldmann Applanation Tonometer

Öz

The aim of this study is to compare the intraocular pressure (IOP) readings taken with the Icare tonometer with the Goldmann applanation tonometer in normal patients. This study involved 64 eyes of 32 patients who was referred to the Ophthalmology clinic of the Uludag University Medical Faculty, for refraction and have no other ophthalmological problem except a refractive error. Two observers obtained IOP readings with Icare tonometer before a local anestetic drop and then also with Goldmann applanation tonometer (GAT). These two readings are compared with each other. For this purpose the analysis of percentage change, intraclass correlation, analysis of Pearson correlation and the Mann- Whitney tests are used; the avarege, median, standart deviation and minimum-maximum values are calculated and the level of significance is accepted as 0.05. In addition, central corneal thickness values of all the patients are obtained with Pentacam®. Also for the assesment of these measurements the Mann Whitney test and analysis of Pearson correlation are used and teh level of significance is also accepted as 0.05. The differance between the two methods (ICare-GAT) determined as for right eyes 0 (ranges -0.40 and 0.82) and for left eyes 0 (ranges between-0.29 and 1.00). There was no significant difference statistically between the IOP measurements of two methods (p> 0.05) The intraclass correlation coefficiency was significantly high with ICare and GAT methods (p

Anahtar Kelimeler

Ayrıntılar

Birincil Dil

İngilizce

Konular

-

Bölüm

-

Yazarlar

Mehmet Baykara Bu kişi benim

Sevil Türüdü Bu kişi benim

Nükhet Zaım Bu kişi benim

Allahyar Ağayarov Bu kişi benim

Hikmet Özçetin Bu kişi benim

Yayımlanma Tarihi

1 Haziran 2011

Gönderilme Tarihi

1 Haziran 2011

Kabul Tarihi

-

Yayımlandığı Sayı

Yıl 2011 Cilt: 37 Sayı: 2

Kaynak Göster

APA
Baykara, M., Türüdü, S., Zaım, N., Ağayarov, A., & Özçetin, H. (2011). Clinical Comparison of Icare Tonometer and Goldmann Applanation Tonometer. Journal of Uludağ University Medical Faculty, 37(2), 89-91. https://izlik.org/JA34FE84TB
AMA
1.Baykara M, Türüdü S, Zaım N, Ağayarov A, Özçetin H. Clinical Comparison of Icare Tonometer and Goldmann Applanation Tonometer. Uludağ Tıp Derg. 2011;37(2):89-91. https://izlik.org/JA34FE84TB
Chicago
Baykara, Mehmet, Sevil Türüdü, Nükhet Zaım, Allahyar Ağayarov, ve Hikmet Özçetin. 2011. “Clinical Comparison of Icare Tonometer and Goldmann Applanation Tonometer”. Journal of Uludağ University Medical Faculty 37 (2): 89-91. https://izlik.org/JA34FE84TB.
EndNote
Baykara M, Türüdü S, Zaım N, Ağayarov A, Özçetin H (01 Haziran 2011) Clinical Comparison of Icare Tonometer and Goldmann Applanation Tonometer. Journal of Uludağ University Medical Faculty 37 2 89–91.
IEEE
[1]M. Baykara, S. Türüdü, N. Zaım, A. Ağayarov, ve H. Özçetin, “Clinical Comparison of Icare Tonometer and Goldmann Applanation Tonometer”, Uludağ Tıp Derg, c. 37, sy 2, ss. 89–91, Haz. 2011, [çevrimiçi]. Erişim adresi: https://izlik.org/JA34FE84TB
ISNAD
Baykara, Mehmet - Türüdü, Sevil - Zaım, Nükhet - Ağayarov, Allahyar - Özçetin, Hikmet. “Clinical Comparison of Icare Tonometer and Goldmann Applanation Tonometer”. Journal of Uludağ University Medical Faculty 37/2 (01 Haziran 2011): 89-91. https://izlik.org/JA34FE84TB.
JAMA
1.Baykara M, Türüdü S, Zaım N, Ağayarov A, Özçetin H. Clinical Comparison of Icare Tonometer and Goldmann Applanation Tonometer. Uludağ Tıp Derg. 2011;37:89–91.
MLA
Baykara, Mehmet, vd. “Clinical Comparison of Icare Tonometer and Goldmann Applanation Tonometer”. Journal of Uludağ University Medical Faculty, c. 37, sy 2, Haziran 2011, ss. 89-91, https://izlik.org/JA34FE84TB.
Vancouver
1.Mehmet Baykara, Sevil Türüdü, Nükhet Zaım, Allahyar Ağayarov, Hikmet Özçetin. Clinical Comparison of Icare Tonometer and Goldmann Applanation Tonometer. Uludağ Tıp Derg [Internet]. 01 Haziran 2011;37(2):89-91. Erişim adresi: https://izlik.org/JA34FE84TB

ISSN: 1300-414X, e-ISSN: 2645-9027

Uludağ Üniversitesi Tıp Fakültesi Dergisi "Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License" ile lisanslanmaktadır.


Creative Commons License
Journal of Uludag University Medical Faculty is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.

2023