Araştırma Makalesi

A Comparison Of Manual Hemocytometry And Different Modes Of Mindray BC-6800 Hematology Analyzer For Cell Count In Peritoneal And Pleural Fluids

Cilt: 47 Sayı: 2 17 Haziran 2020
  • Ünsal Savcı *
  • Mustafa Şahin
  • Barış Eser
PDF İndir

A Comparison Of Manual Hemocytometry And Different Modes Of Mindray BC-6800 Hematology Analyzer For Cell Count In Peritoneal And Pleural Fluids

Öz

Objective: Total white blood cell (WBC) count in body fluids (BF) is important in diagnosis and treatment of inflammatory and infectious diseases. We aimed to compare hemogram and BF modes of Mindray BC-6800 hematology analyzer with the manual hemocytometry method for determination of total WBC count in peritoneal and pleural fluids. Method: Our study consisted of a total of 143 specimens, 109 peritoneal fluid and 34 pleural effusion fluid. Each sample was analyzed consecutively twice, first with the manual method and, later with both hemogram mode and the BF mode of Mindray BC-6800 automated hematology analyzer. Results: There was a statistically significant difference between manual count and hemogram mode (for peritoneal fluids p< 0.001, for pleural fluids p< 0.001). There was a statistically significant difference between the hemogram mode and BF mode too. (for peritoneal fluids p< 0.001, for pleural fluids p< 0.001). However, there was no statistically significant difference between the manual count and BF mode (for peritoneal fluids p=0.236, for pleural fluids p=0.627). Less than100 cells/mL, there was a statistically significant difference between each of the counting methods (Manual count vs hemogram mode p< 0.001, manual count vs BF mode p= 0.012, hemogram mode vs BF mode p< 0.001). More than 100 cells/mL, there was no statistically significant difference between manual count and the BF mode (p= 0.332). However, there were statistically significant differences between manual count and hemogram mode, and hemogram mode and BF mode (p= 0.003 and p< 0.001 respectively). Conclusion: The use of manual method instead of hematology analyzer is more convenient in samples <100 cells/mL, and BF mode may be used as a screening tool in samples >100 cells/mL.

Anahtar Kelimeler

Kaynakça

  1. 1. De Smet D, Van Moer G, Martens GA, et al. Use of the Cell-Dyn Sapphire hematology analyzer for automated counting of blood cells in body fluids. Am J Clin Pathol. 2010; 133: 291-9.
  2. 2. Cognialli RCR, Comar SR, Souza AMD, Singer GMB. Evaluation of pleural and ascitic fluid analysis on the Sysmex XE-5000 hematology analyzer. J Bras Patol Med Lab. 2017; 53; 150-8.
  3. 3. Fuster O, Andino B, Pardo A, Laiz B. Continuous ambulatory peritoneal dialysis, ascitic and pleural body fluids evaluation with the Mindray BC-6800 hematology analyzer. Journal of clinical laboratory analysis. J Clin Lab Anal. 2018; 32: e22240.
  4. 4. Runyon BA. Management of adult patients with ascites due to cirrhosis: an update. Hepatology. 2009; 49: 2087-107.
  5. 5. Conner BD, Lee YG, Branca P, Rogers JT, Rodriguez RM, Light RW. Variations in pleural fluid WBC count and differential counts with different sample containers and different methods. Chest. 2003; 123: 1181-7.
  6. 6. Bottini PV, Garlipp CR, Nogueira BF, Pompeo DB, Souza MI (2018). Performance evaluation of a new generation of automated analyzer for pleural and peritoneal fluids cytology. Clin Lab. 2018; 64: 1783-6.
  7. 7. Schonenberg H, Fassl H. Comparison of the coulter-counter-method and the counting-chamber to determine the cell count of the cerebrospinal fluid. Arztliche Forschung. 1970; 24: 118.
  8. 8. Walker TJ, Nelson LD, Dunphy BW, Anderson DM, Kickler TS. Comparative Evaluation of the Iris iQ200 Body Fluid Module With Manual Hemacytometer Count. Am J Clin Pathol. 2009; 131: 333-8.

Ayrıntılar

Birincil Dil

İngilizce

Konular

Sağlık Kurumları Yönetimi

Bölüm

Araştırma Makalesi

Yazarlar

Ünsal Savcı * Bu kişi benim
Türkiye

Mustafa Şahin Bu kişi benim
Türkiye

Barış Eser Bu kişi benim
Türkiye

Yayımlanma Tarihi

17 Haziran 2020

Gönderilme Tarihi

24 Eylül 2019

Kabul Tarihi

10 Mayıs 2020

Yayımlandığı Sayı

Yıl 2020 Cilt: 47 Sayı: 2

Kaynak Göster

APA
Savcı, Ü., Şahin, M., & Eser, B. (2020). A Comparison Of Manual Hemocytometry And Different Modes Of Mindray BC-6800 Hematology Analyzer For Cell Count In Peritoneal And Pleural Fluids. Dicle Medical Journal, 47(2), 278-285. https://doi.org/10.5798/dicletip.755702
AMA
1.Savcı Ü, Şahin M, Eser B. A Comparison Of Manual Hemocytometry And Different Modes Of Mindray BC-6800 Hematology Analyzer For Cell Count In Peritoneal And Pleural Fluids. diclemedj. 2020;47(2):278-285. doi:10.5798/dicletip.755702
Chicago
Savcı, Ünsal, Mustafa Şahin, ve Barış Eser. 2020. “A Comparison Of Manual Hemocytometry And Different Modes Of Mindray BC-6800 Hematology Analyzer For Cell Count In Peritoneal And Pleural Fluids”. Dicle Medical Journal 47 (2): 278-85. https://doi.org/10.5798/dicletip.755702.
EndNote
Savcı Ü, Şahin M, Eser B (01 Haziran 2020) A Comparison Of Manual Hemocytometry And Different Modes Of Mindray BC-6800 Hematology Analyzer For Cell Count In Peritoneal And Pleural Fluids. Dicle Medical Journal 47 2 278–285.
IEEE
[1]Ü. Savcı, M. Şahin, ve B. Eser, “A Comparison Of Manual Hemocytometry And Different Modes Of Mindray BC-6800 Hematology Analyzer For Cell Count In Peritoneal And Pleural Fluids”, diclemedj, c. 47, sy 2, ss. 278–285, Haz. 2020, doi: 10.5798/dicletip.755702.
ISNAD
Savcı, Ünsal - Şahin, Mustafa - Eser, Barış. “A Comparison Of Manual Hemocytometry And Different Modes Of Mindray BC-6800 Hematology Analyzer For Cell Count In Peritoneal And Pleural Fluids”. Dicle Medical Journal 47/2 (01 Haziran 2020): 278-285. https://doi.org/10.5798/dicletip.755702.
JAMA
1.Savcı Ü, Şahin M, Eser B. A Comparison Of Manual Hemocytometry And Different Modes Of Mindray BC-6800 Hematology Analyzer For Cell Count In Peritoneal And Pleural Fluids. diclemedj. 2020;47:278–285.
MLA
Savcı, Ünsal, vd. “A Comparison Of Manual Hemocytometry And Different Modes Of Mindray BC-6800 Hematology Analyzer For Cell Count In Peritoneal And Pleural Fluids”. Dicle Medical Journal, c. 47, sy 2, Haziran 2020, ss. 278-85, doi:10.5798/dicletip.755702.
Vancouver
1.Ünsal Savcı, Mustafa Şahin, Barış Eser. A Comparison Of Manual Hemocytometry And Different Modes Of Mindray BC-6800 Hematology Analyzer For Cell Count In Peritoneal And Pleural Fluids. diclemedj. 01 Haziran 2020;47(2):278-85. doi:10.5798/dicletip.755702