Evaluation of the Concordance of Manual Microscopy Results with Automatic Hematology Analyzer in Cell Counting in Body Fluids

Yıl 2026, Cilt: 27 Sayı: 2 , 182 - 187 , 16.04.2026
https://doi.org/10.18229/kocatepetip.1671777
https://izlik.org/JA75CF23HA

Öz

OBJECTIVE: In this study, it was aimed to evaluate the compatibility of white blood cell (WBC) count results obtained using an automatic blood count analyzer with the results obtained by manual microscopy, which is the reference method.
MATERIAL AND METHODS: Between 2019 and 2020, sterile body fluid samples sent simultaneously to the medical biochemistry laboratory for automated body fluid cell count and to the medical microbiology laboratory for manual cell count were examined. During the manual microscopy stage, the samples were evaluated by looking at the microscope with a Thoma slide by two different researchers. WBC measurement was performed using the body fluid module (Sysmex Corporation, Kobe, Japan) on the Sysmex XN-10 blood counting analyzer.
RESULTS: In our study, a total of 79 sterile body fluids (40 synovial fluid, 21 pleural fluid, 14 peritoneal fluid, 4 cerebrospinal fluid) were analyzed. "Median (25%-75%) value was determined as 1520 (60-8340) cells/μL according to manual white blood cell count and 1615 (68-6091) cells/μL according to Sysmex XN-10. A statistically significant strong correlation was found between Sysmex XN-10 and manual microscopy. Statistical comparisons conducted with Passing-Bablok regression and Bland-Altman regression analyses revealed good agreement between manual white blood cell count and XN-10. Based on the 500 cells/μL cut-off value, the automated hematology analyzer demonstrated high diagnostic performance (sensitivity 93.9%, specificity 87.1%). The PPV, NPV, and overall accuracy were calculated as 92.0%, 90.0%, and 91.2%, respectively.
CONCLUSIONS: Cell counting in sterile body fluids using an automatic analyzer is becoming increasingly common. In our study, it was determined that body samples gave generally compatible results with manual microscopy.

Kaynakça

  • 1. Seghezzi M, Previtali G, Moioli V, et al. Performance evaluation of automated cell counts compared with reference methods for body fluid analysis. Diagnosis (Berl) 2021;9(3):369-78.
  • 2. Buoro S, Seghezzi M, Dominoni P, Moioli V, et al. Lack of harmonization in high fluorescent cell automated counts with body fluids mode in ascitic, pleural, synovial, and cerebrospinal fluids. Int J Lab Hematol 2019;41(2):277-86.
  • 3. Aguadero V, Cano-Corres R, Berlanga E, et al. Evaluation of biological fluid analysis using the Sysmex XN automatic hematology analyzer. Cytometry B Clin Cytom 2018;94(5):680-8.
  • 4. Sandhaus LM. Body fluid cell counts by automated methods. Clin Lab Med 2015;35(1):93-103.
  • 5. Andrews J, Setran E, McDonnel L, et al. An evaluation of the cell-dyn 3200 for counting cells in cerebrospinal and other body fluids. Lab Hematol 2005;11(2):98-106.
  • 6. Cho J, Oh J, Lee S, et al. Performance evaluation of body fluid cellular analysis using the Beckman Coulter UniCel DxH 800, Sysmex XN-350, and UF-5000 automated cellular analyzers. Ann Lab Med 2020;40:122-30.
  • 7. Dancey CP, Reidy J. Statistics Without Maths for Psychology. 8th Edition, Pearson Education. 2020:145-56.
  • 8. Fleming C, Russcher H, Lindemans J, et al. Clinical relevance and contemporary methods for counting blood cells in body fluids suspected of inflammatory disease. Clin Chem Lab Med 2015;53:1689-706.
  • 9. Fındık D. Steril Vücut Sıvıları Laboratuvar İnceleme Rehberi, KLİMUD, 2. Baskı. https://www.klimud.org/uploads/content/Steril%20vucut%20sivilari_ver02_yeni%20dizgi%20291020_1.pdf, Erişim Tarihi: 20.04.2025
  • 10. Ivady G, Barath S, Szaraz-Szeles M, et al. Comparative Evaluation of body fluid analysis by Sysmex XN hematology analyzers, CellaVision, manual microscopy and multicolor flow cytometry. Ann Clin Lab Sci 2022;52(2):314-22.
  • 11. Evren E, Us E, Karahan ZC. Determining white blood cell count in sterile body fluids: cell counting chamber vs Sysmex UF-1000i. GMJ 2020;31:19-21.
  • 12. Paris A, Nhan T, Cornet E, et al. Performance evaluation of the body fluid mode on the platform Sysmex XE- 5000 series automated hematology analyzer. Int J Lab Hematol 2010;32(5):539-47.
  • 13. Danise P, Maconi M, Rovetti A, et al. Cell counting of body fluids: comparison between three automated haematology analysers and the manual microscope method. Int J Lab Hematol 2013;35(6):608-13.
  • 14. Glasser L, Murphy CA, Machan JT. The clinical reliability of automated cerebrospinal fluid cell counts on the Beckman-Coulter LH750 and Iris iQ200. Am J Clin Pathol 2009;131(1):58-63.
  • 15. Heller T, Nagel I, Ehrlich B, et al. Automated cerebrospinal fluid cytology. Anal Quant Cytol Histol 2008;30(3):139-44.
  • 16. Siatkowski M, Dahyot S, Pestel-Caron M, et al. Performance evaluation of UF-4000 body fluid mode for automated body fluid cell counting. Clin Chim Acta 2022;531:152-6.
  • 17. Baran E, Ilki AA. Evaluation of Sysmex-UF 5000-BF Module for sterile body fluids. An alternative for conventional methods? Clin Lab 2023;69(4).
  • 18. Boer K, Deufel T, Reinhoefer M. Evaluation of the XE-5000 for the automated analysis of blood cells in cerebrospinal fluid. Clin Biochem 2009;42:684-91.
  • 19. Buoro S, Apassiti Esposito S, Alessio M, et al. Automated cerebrospinal fluid cell counts using the new body fluid mode of Sysmex UF-1000i. J Clin Lab Anal 2016;30:381-91.
  • 20. Alcaide Martín MJ, Altimira Queral L, Sahuquillo Frías L, et al. Automated cell count in body fluids: a review. Adv Lab Med 2021;2(2):149-77.
  • 21. American Academy of Pediatrics. Red Book: 2021–2024 Report of the Committee on Infectious Diseases. 32nd ed. Elk Grove Village (IL): American Academy of Pediatrics; 2021;(32): 572–6.

VÜCUT SIVISI HÜCRE SAYIMINDA OTOMATİK HEMATOLOJİ ANALİZÖRÜ İLE MANUEL MİKROSKOPİ SONUÇLARININ UYUMUNUN DEĞERLENDİRİLMESİ

Yıl 2026, Cilt: 27 Sayı: 2 , 182 - 187 , 16.04.2026
https://doi.org/10.18229/kocatepetip.1671777
https://izlik.org/JA75CF23HA

Öz

AMAÇ: Bu çalışmada otomatik kan sayım analizörü kullanılarak elde edilen beyaz küre hücre (WBC) sayım sonuçlarının referans yöntem olan manuel mikroskopi ile elde edilen sonuçlarla uyumunun değerlendirilmesi amaçlanmıştır.
GEREÇ VE YÖNTEM: 2019-2020 yılları arasında tıbbi biyokimya laboratuvarına otomatize vücut sıvısı hücre sayımı, tıbbi mikrobiyoloji laboratuvarına manuel hücre sayımı yapılması amacıyla gönderilen steril vücut sıvısı örnekleri retrospektif olarak incelendi. Manuel mikroskopi aşamasında örnekler iki farklı araştırmacı tarafından Thoma lamı ile ışık mikroskopunda bakılarak değerlendirildi. Vücut sıvısının otomatize WBC ölçümü Sysmex XN-10 kan sayım cihazında vücut sıvısı modülü (Sysmex Corporation, Kobe, Japonya) kullanılarak analiz yapıldı.
BULGULAR: Çalışmamızda toplam 79 steril vücut sıvısı (40 sinovyal sıvı, 21 plevra sıvı, 14 periton sıvısı, 4 beyin omurilik sıvısı) analiz edildi. Manuel beyaz küre hücre sayımına göre medyan (25%-75%) değeri 1520 (60-8340) hücre/μL, Sysmex XN-10'a göre ise 1615 (68-6091) hücre/μL olarak saptandı. XN-10 ve manuel mikroskopi arasında istatistiksel olarak anlamlı güçlü bir korelasyon tespit edildi. Passing-Bablok regresyon ve Bland-Altman farklar analizi ile yapılan istatistiksel karşılaştırmalarda manuel beyaz küre hücre sayımı ve XN-10 arasında iyi bir uyum tespit edildi. 500 hücre/μL eşik değeriyle yapılan değerlendirmede, otomatik analizör yüksek tanısal performans göstermiştir (duyarlılık %93,9, özgüllük %87,1). PPV %92,0, NPV %90,0 ve doğruluk %91,2 olarak hesaplanmıştır.
SONUÇ: Steril vücut sıvılarında hücre sayımının otomatik cihaz ile yapılması giderek yaygınlaşmaktadır. Yaptığımız çalışmada vücut örneklerinde manuel mikroskopiyle genel olarak uyumlu sonuçlar verdiği saptandı.

Kaynakça

  • 1. Seghezzi M, Previtali G, Moioli V, et al. Performance evaluation of automated cell counts compared with reference methods for body fluid analysis. Diagnosis (Berl) 2021;9(3):369-78.
  • 2. Buoro S, Seghezzi M, Dominoni P, Moioli V, et al. Lack of harmonization in high fluorescent cell automated counts with body fluids mode in ascitic, pleural, synovial, and cerebrospinal fluids. Int J Lab Hematol 2019;41(2):277-86.
  • 3. Aguadero V, Cano-Corres R, Berlanga E, et al. Evaluation of biological fluid analysis using the Sysmex XN automatic hematology analyzer. Cytometry B Clin Cytom 2018;94(5):680-8.
  • 4. Sandhaus LM. Body fluid cell counts by automated methods. Clin Lab Med 2015;35(1):93-103.
  • 5. Andrews J, Setran E, McDonnel L, et al. An evaluation of the cell-dyn 3200 for counting cells in cerebrospinal and other body fluids. Lab Hematol 2005;11(2):98-106.
  • 6. Cho J, Oh J, Lee S, et al. Performance evaluation of body fluid cellular analysis using the Beckman Coulter UniCel DxH 800, Sysmex XN-350, and UF-5000 automated cellular analyzers. Ann Lab Med 2020;40:122-30.
  • 7. Dancey CP, Reidy J. Statistics Without Maths for Psychology. 8th Edition, Pearson Education. 2020:145-56.
  • 8. Fleming C, Russcher H, Lindemans J, et al. Clinical relevance and contemporary methods for counting blood cells in body fluids suspected of inflammatory disease. Clin Chem Lab Med 2015;53:1689-706.
  • 9. Fındık D. Steril Vücut Sıvıları Laboratuvar İnceleme Rehberi, KLİMUD, 2. Baskı. https://www.klimud.org/uploads/content/Steril%20vucut%20sivilari_ver02_yeni%20dizgi%20291020_1.pdf, Erişim Tarihi: 20.04.2025
  • 10. Ivady G, Barath S, Szaraz-Szeles M, et al. Comparative Evaluation of body fluid analysis by Sysmex XN hematology analyzers, CellaVision, manual microscopy and multicolor flow cytometry. Ann Clin Lab Sci 2022;52(2):314-22.
  • 11. Evren E, Us E, Karahan ZC. Determining white blood cell count in sterile body fluids: cell counting chamber vs Sysmex UF-1000i. GMJ 2020;31:19-21.
  • 12. Paris A, Nhan T, Cornet E, et al. Performance evaluation of the body fluid mode on the platform Sysmex XE- 5000 series automated hematology analyzer. Int J Lab Hematol 2010;32(5):539-47.
  • 13. Danise P, Maconi M, Rovetti A, et al. Cell counting of body fluids: comparison between three automated haematology analysers and the manual microscope method. Int J Lab Hematol 2013;35(6):608-13.
  • 14. Glasser L, Murphy CA, Machan JT. The clinical reliability of automated cerebrospinal fluid cell counts on the Beckman-Coulter LH750 and Iris iQ200. Am J Clin Pathol 2009;131(1):58-63.
  • 15. Heller T, Nagel I, Ehrlich B, et al. Automated cerebrospinal fluid cytology. Anal Quant Cytol Histol 2008;30(3):139-44.
  • 16. Siatkowski M, Dahyot S, Pestel-Caron M, et al. Performance evaluation of UF-4000 body fluid mode for automated body fluid cell counting. Clin Chim Acta 2022;531:152-6.
  • 17. Baran E, Ilki AA. Evaluation of Sysmex-UF 5000-BF Module for sterile body fluids. An alternative for conventional methods? Clin Lab 2023;69(4).
  • 18. Boer K, Deufel T, Reinhoefer M. Evaluation of the XE-5000 for the automated analysis of blood cells in cerebrospinal fluid. Clin Biochem 2009;42:684-91.
  • 19. Buoro S, Apassiti Esposito S, Alessio M, et al. Automated cerebrospinal fluid cell counts using the new body fluid mode of Sysmex UF-1000i. J Clin Lab Anal 2016;30:381-91.
  • 20. Alcaide Martín MJ, Altimira Queral L, Sahuquillo Frías L, et al. Automated cell count in body fluids: a review. Adv Lab Med 2021;2(2):149-77.
  • 21. American Academy of Pediatrics. Red Book: 2021–2024 Report of the Committee on Infectious Diseases. 32nd ed. Elk Grove Village (IL): American Academy of Pediatrics; 2021;(32): 572–6.
Toplam 21 adet kaynakça vardır.

Ayrıntılar

Birincil Dil Türkçe
Konular Klinik Mikrobiyoloji, Tıbbi Biyokimya ve Metabolomik (Diğer)
Bölüm Araştırma Makalesi
Yazarlar

Nihan Ceken 0000-0003-1877-7320

Medine Alpdemir 0000-0003-2625-0246

Hülya Duran 0000-0002-4838-0730

Mehmet Fatih Alpdemir 0000-0003-3015-1478

Tuğba Kula Atik 0000-0002-2433-1977

Gönderilme Tarihi 8 Nisan 2025
Kabul Tarihi 21 Ağustos 2025
Yayımlanma Tarihi 16 Nisan 2026
DOI https://doi.org/10.18229/kocatepetip.1671777
IZ https://izlik.org/JA75CF23HA
Yayımlandığı Sayı Yıl 2026 Cilt: 27 Sayı: 2

Kaynak Göster

APA Ceken, N., Alpdemir, M., Duran, H., Alpdemir, M. F., & Kula Atik, T. (2026). VÜCUT SIVISI HÜCRE SAYIMINDA OTOMATİK HEMATOLOJİ ANALİZÖRÜ İLE MANUEL MİKROSKOPİ SONUÇLARININ UYUMUNUN DEĞERLENDİRİLMESİ. Kocatepe Tıp Dergisi, 27(2), 182-187. https://doi.org/10.18229/kocatepetip.1671777
AMA 1.Ceken N, Alpdemir M, Duran H, Alpdemir MF, Kula Atik T. VÜCUT SIVISI HÜCRE SAYIMINDA OTOMATİK HEMATOLOJİ ANALİZÖRÜ İLE MANUEL MİKROSKOPİ SONUÇLARININ UYUMUNUN DEĞERLENDİRİLMESİ. KTD. 2026;27(2):182-187. doi:10.18229/kocatepetip.1671777
Chicago Ceken, Nihan, Medine Alpdemir, Hülya Duran, Mehmet Fatih Alpdemir, ve Tuğba Kula Atik. 2026. “VÜCUT SIVISI HÜCRE SAYIMINDA OTOMATİK HEMATOLOJİ ANALİZÖRÜ İLE MANUEL MİKROSKOPİ SONUÇLARININ UYUMUNUN DEĞERLENDİRİLMESİ”. Kocatepe Tıp Dergisi 27 (2): 182-87. https://doi.org/10.18229/kocatepetip.1671777.
EndNote Ceken N, Alpdemir M, Duran H, Alpdemir MF, Kula Atik T (01 Nisan 2026) VÜCUT SIVISI HÜCRE SAYIMINDA OTOMATİK HEMATOLOJİ ANALİZÖRÜ İLE MANUEL MİKROSKOPİ SONUÇLARININ UYUMUNUN DEĞERLENDİRİLMESİ. Kocatepe Tıp Dergisi 27 2 182–187.
IEEE [1]N. Ceken, M. Alpdemir, H. Duran, M. F. Alpdemir, ve T. Kula Atik, “VÜCUT SIVISI HÜCRE SAYIMINDA OTOMATİK HEMATOLOJİ ANALİZÖRÜ İLE MANUEL MİKROSKOPİ SONUÇLARININ UYUMUNUN DEĞERLENDİRİLMESİ”, KTD, c. 27, sy 2, ss. 182–187, Nis. 2026, doi: 10.18229/kocatepetip.1671777.
ISNAD Ceken, Nihan - Alpdemir, Medine - Duran, Hülya - Alpdemir, Mehmet Fatih - Kula Atik, Tuğba. “VÜCUT SIVISI HÜCRE SAYIMINDA OTOMATİK HEMATOLOJİ ANALİZÖRÜ İLE MANUEL MİKROSKOPİ SONUÇLARININ UYUMUNUN DEĞERLENDİRİLMESİ”. Kocatepe Tıp Dergisi 27/2 (01 Nisan 2026): 182-187. https://doi.org/10.18229/kocatepetip.1671777.
JAMA 1.Ceken N, Alpdemir M, Duran H, Alpdemir MF, Kula Atik T. VÜCUT SIVISI HÜCRE SAYIMINDA OTOMATİK HEMATOLOJİ ANALİZÖRÜ İLE MANUEL MİKROSKOPİ SONUÇLARININ UYUMUNUN DEĞERLENDİRİLMESİ. KTD. 2026;27:182–187.
MLA Ceken, Nihan, vd. “VÜCUT SIVISI HÜCRE SAYIMINDA OTOMATİK HEMATOLOJİ ANALİZÖRÜ İLE MANUEL MİKROSKOPİ SONUÇLARININ UYUMUNUN DEĞERLENDİRİLMESİ”. Kocatepe Tıp Dergisi, c. 27, sy 2, Nisan 2026, ss. 182-7, doi:10.18229/kocatepetip.1671777.
Vancouver 1.Nihan Ceken, Medine Alpdemir, Hülya Duran, Mehmet Fatih Alpdemir, Tuğba Kula Atik. VÜCUT SIVISI HÜCRE SAYIMINDA OTOMATİK HEMATOLOJİ ANALİZÖRÜ İLE MANUEL MİKROSKOPİ SONUÇLARININ UYUMUNUN DEĞERLENDİRİLMESİ. KTD. 01 Nisan 2026;27(2):182-7. doi:10.18229/kocatepetip.1671777

88x31.png
Bu Dergi Creative Commons Atıf-GayriTicari-AynıLisanslaPaylaş 4.0 Uluslararası Lisansı ile lisanslanmıştır.