Araştırma Makalesi
BibTex RIS Kaynak Göster

Radiation Dose Measurements around Patient in Computed Tomography Imaging

Yıl 2019, Cilt: 23 Sayı: 3, 792 - 796, 25.12.2019
https://doi.org/10.19113/sdufenbed.544773

Öz

Ionizing
radiation is intensively used in diagnostic medical imaging and Computed
tomography (CT) is the most requested among the modalities. During exposure,
X-rays are usually scattered from the patient and the system according to the
radiation physics laws. Thus, estimation of the received dose caused by the
scattered radiation is important for the patients, patients' companions and the
healthcare workers. The aim of this study was to determine the radiation dose
from the patient’s body and various materials in the environment during CT
imaging. In this study, chest and head and neck CT scans were performed on
human tissue equivalent Alderson Rando phantom. During CT imaging, radiation
dose measurements were achieved by thermoluminescent dosimeters (TLD) placed at
distances of 10, 20, 30 and 40 cm from the phantom. In chest CT imaging, the
mean radiation dose to the environment ranged from 6.75 ± 1.07 µSv to 21.68 ±
1.45 µSv. While, Head and neck imaging led to radiation dose ranged from 8.38 ±
0.81 µSv to 26.57 ± 0.98 µSv. The exposure danger of the accompanying
individuals was found to be minimal and below the permissible limits.

Kaynakça

  • [1] Sodickson, A., Baeyens, P.F., Andriole, K.P., Prevedello, L.M., Nawfel, R.D., Hanson, R., Khorasani, R., 2009. Recurrent CT, cumulative radiation exposure, and associated radiation-induced cancer risks from CT of adults. Radiology, 251(1), 175-184.
  • [2] Cherry SR, Sorenson JA Phelps ME.,2003, Physics in Nuclear Medicine, 3rd edition, Philadephia.
  • [3] McNitt-Gray, M.F., 2002. AAPM/RSNA physics tutorial for residents: topics in CT: radiation dose in CT. Radiographics, 22(6), 1541-1553.
  • [4] Işık Z, Selçuk H, Albayram S.,2010, Bilgisayarlı Tomografi ve Radyasyon. Klinik Gelişim, 23, 16-18.
  • [5] Valentin, J., 2007. The 2007 recommendations of the international commission on radiological protection. ICRP publication 103. Ann iCRP, 37(2), 1-332.
  • [6] Allisy-Roberts, P.J., Williams, J., 2007. Farr's physics for medical imaging. Elsevier Health Sciences.
  • [7] Wildgruber, M., Müller-Wille, R., Goessmann, H., Uller, W., Wohlgemuth, W.A., 2016. Direct effective dose calculations in pediatric fluoroscopy-guided abdominal interventions with rando-alderson phantoms–optimization of preset parameter settings. PloS one, 11(8).
  • [8] Lee, G.S., Kim, J.S., Seo, Y.S., Kim, J.D., 2013. Effective dose from direct and indirect digital panoramic units. Imaging science in dentistry, 43(2), 77-84.
  • [9] Harshaw-Bicron, 1992. TLD Radiation Evaluation and Management System(TLD-REMS) User’s Manual for use with TLD 8800 & 6600 Card Readers.REMS-0-U-0492-006. Bicron, Saint-Gobain/Norton Industrial CeramicsCorporation, Solon, OH, USA.
  • [10] Harshaw-Bicron, 1994. Model 6600E Automatic TLD Workstation User’sManual. Publication no. 6600-E-U-0294-001. Bicron, Saint-Gobain/NortonIndustrial Ceramics Corporation, Solon, OH, USA7
  • [11] Tekin, H.O., Manici, T., Ekmekci, C., 2016. Investigation of backscattered dose in a computerized tomography (CT) facility during abdominal CT scan by considering clinical measurements and application of Monte Carlo method. Journal of Health Science, 4, 131-134.
  • [12] Tekin H. O., Cavlı, B., Altunsoy, E. E., Manici, T., Ozturk, C., Karakas, H. Ml., 2018, An investigation on radiation protection and shielding properties of 16 slice computed tomography (CT) facilities. International Journal of Computational and Experimental Science and Engineering, 4(2), 37-40.

Bilgisayarlı Tomografi Çekimlerinde Hastanın Yakın Çevresinde Radyasyon Dozu Ölçümleri

Yıl 2019, Cilt: 23 Sayı: 3, 792 - 796, 25.12.2019
https://doi.org/10.19113/sdufenbed.544773

Öz

İyonize
radyasyonun en yoğun kullanıldığı görüntüleme yöntemlerinden biri bilgisayarlı
tomografi (BT) dir. Bilgisayarlı tomografi çekimlerinde X-ışınları kullanılır.
Çekim sırasında X-ışınlarının bir kısmı, radyasyon fiziği yasalarına göre
hastadan ve sistemden çevreye saçılır. Bu saçılan radyasyondan kaynaklanan radyasyon
dozunun belirlenmesi hasta, hastaya eşlik etme zorunda kalan hasta yakını ve sağlık
çalışanları açısından önemlidir. Bu çalışmanın amacı BT görüntülemesi sırasında
hastadan ve ortamdaki çeşitli materyallerden çevreye yayılan radyasyondan
kaynaklanan dozunun belirlenmesidir. Bu çalışmada insan eşdeğeri olan Alderson
Rando fantomun göğüs ve baş-boyun BT görüntülemesi yapıldı. BT görüntüleme
sırasında fantomdan 10, 20, 30 ve 40 cm uzaklıklara termolüminisans
dozimetreler (TLD) yerleştirilerek radyasyon dozu ölçümleri yapıldı TLD
dozimetrelerin kalibrasyonları ve okumaları Çekmece Nükleer Araştırmalar
Merkezinde yapıldı. Göğüs BT görüntülemede çevreye saçılan ortalama radyasyon
dozunun
6.75±1.07
µSv
ile
21.68±1.45 µSv arasında değiştiği belirlendi. Baş-boyun
görüntülemede ise çevreye saçılan ortalama radyasyon dozunun
8.38±0.81 µSv ile 26.57±0.98 µSv
arasında değiştiği belirlendi. Çekim sırasında hastaya eşlik etmek zorunda
kalan şahısların doz maruziyetlerinin müsaade edilen limitlerin altında olduğu tespit
edildi.

Kaynakça

  • [1] Sodickson, A., Baeyens, P.F., Andriole, K.P., Prevedello, L.M., Nawfel, R.D., Hanson, R., Khorasani, R., 2009. Recurrent CT, cumulative radiation exposure, and associated radiation-induced cancer risks from CT of adults. Radiology, 251(1), 175-184.
  • [2] Cherry SR, Sorenson JA Phelps ME.,2003, Physics in Nuclear Medicine, 3rd edition, Philadephia.
  • [3] McNitt-Gray, M.F., 2002. AAPM/RSNA physics tutorial for residents: topics in CT: radiation dose in CT. Radiographics, 22(6), 1541-1553.
  • [4] Işık Z, Selçuk H, Albayram S.,2010, Bilgisayarlı Tomografi ve Radyasyon. Klinik Gelişim, 23, 16-18.
  • [5] Valentin, J., 2007. The 2007 recommendations of the international commission on radiological protection. ICRP publication 103. Ann iCRP, 37(2), 1-332.
  • [6] Allisy-Roberts, P.J., Williams, J., 2007. Farr's physics for medical imaging. Elsevier Health Sciences.
  • [7] Wildgruber, M., Müller-Wille, R., Goessmann, H., Uller, W., Wohlgemuth, W.A., 2016. Direct effective dose calculations in pediatric fluoroscopy-guided abdominal interventions with rando-alderson phantoms–optimization of preset parameter settings. PloS one, 11(8).
  • [8] Lee, G.S., Kim, J.S., Seo, Y.S., Kim, J.D., 2013. Effective dose from direct and indirect digital panoramic units. Imaging science in dentistry, 43(2), 77-84.
  • [9] Harshaw-Bicron, 1992. TLD Radiation Evaluation and Management System(TLD-REMS) User’s Manual for use with TLD 8800 & 6600 Card Readers.REMS-0-U-0492-006. Bicron, Saint-Gobain/Norton Industrial CeramicsCorporation, Solon, OH, USA.
  • [10] Harshaw-Bicron, 1994. Model 6600E Automatic TLD Workstation User’sManual. Publication no. 6600-E-U-0294-001. Bicron, Saint-Gobain/NortonIndustrial Ceramics Corporation, Solon, OH, USA7
  • [11] Tekin, H.O., Manici, T., Ekmekci, C., 2016. Investigation of backscattered dose in a computerized tomography (CT) facility during abdominal CT scan by considering clinical measurements and application of Monte Carlo method. Journal of Health Science, 4, 131-134.
  • [12] Tekin H. O., Cavlı, B., Altunsoy, E. E., Manici, T., Ozturk, C., Karakas, H. Ml., 2018, An investigation on radiation protection and shielding properties of 16 slice computed tomography (CT) facilities. International Journal of Computational and Experimental Science and Engineering, 4(2), 37-40.
Toplam 12 adet kaynakça vardır.

Ayrıntılar

Birincil Dil Türkçe
Konular Mühendislik
Bölüm Makaleler
Yazarlar

Osman Günay 0000-0003-0760-554X

Mustafa Demir 0000-0002-9813-1628

Yayımlanma Tarihi 25 Aralık 2019
Yayımlandığı Sayı Yıl 2019 Cilt: 23 Sayı: 3

Kaynak Göster

APA Günay, O., & Demir, M. (2019). Bilgisayarlı Tomografi Çekimlerinde Hastanın Yakın Çevresinde Radyasyon Dozu Ölçümleri. Süleyman Demirel Üniversitesi Fen Bilimleri Enstitüsü Dergisi, 23(3), 792-796. https://doi.org/10.19113/sdufenbed.544773
AMA Günay O, Demir M. Bilgisayarlı Tomografi Çekimlerinde Hastanın Yakın Çevresinde Radyasyon Dozu Ölçümleri. Süleyman Demirel Üniv. Fen Bilim. Enst. Derg. Aralık 2019;23(3):792-796. doi:10.19113/sdufenbed.544773
Chicago Günay, Osman, ve Mustafa Demir. “Bilgisayarlı Tomografi Çekimlerinde Hastanın Yakın Çevresinde Radyasyon Dozu Ölçümleri”. Süleyman Demirel Üniversitesi Fen Bilimleri Enstitüsü Dergisi 23, sy. 3 (Aralık 2019): 792-96. https://doi.org/10.19113/sdufenbed.544773.
EndNote Günay O, Demir M (01 Aralık 2019) Bilgisayarlı Tomografi Çekimlerinde Hastanın Yakın Çevresinde Radyasyon Dozu Ölçümleri. Süleyman Demirel Üniversitesi Fen Bilimleri Enstitüsü Dergisi 23 3 792–796.
IEEE O. Günay ve M. Demir, “Bilgisayarlı Tomografi Çekimlerinde Hastanın Yakın Çevresinde Radyasyon Dozu Ölçümleri”, Süleyman Demirel Üniv. Fen Bilim. Enst. Derg., c. 23, sy. 3, ss. 792–796, 2019, doi: 10.19113/sdufenbed.544773.
ISNAD Günay, Osman - Demir, Mustafa. “Bilgisayarlı Tomografi Çekimlerinde Hastanın Yakın Çevresinde Radyasyon Dozu Ölçümleri”. Süleyman Demirel Üniversitesi Fen Bilimleri Enstitüsü Dergisi 23/3 (Aralık 2019), 792-796. https://doi.org/10.19113/sdufenbed.544773.
JAMA Günay O, Demir M. Bilgisayarlı Tomografi Çekimlerinde Hastanın Yakın Çevresinde Radyasyon Dozu Ölçümleri. Süleyman Demirel Üniv. Fen Bilim. Enst. Derg. 2019;23:792–796.
MLA Günay, Osman ve Mustafa Demir. “Bilgisayarlı Tomografi Çekimlerinde Hastanın Yakın Çevresinde Radyasyon Dozu Ölçümleri”. Süleyman Demirel Üniversitesi Fen Bilimleri Enstitüsü Dergisi, c. 23, sy. 3, 2019, ss. 792-6, doi:10.19113/sdufenbed.544773.
Vancouver Günay O, Demir M. Bilgisayarlı Tomografi Çekimlerinde Hastanın Yakın Çevresinde Radyasyon Dozu Ölçümleri. Süleyman Demirel Üniv. Fen Bilim. Enst. Derg. 2019;23(3):792-6.

e-ISSN :1308-6529
Linking ISSN (ISSN-L): 1300-7688

Dergide yayımlanan tüm makalelere ücretiz olarak erişilebilinir ve Creative Commons CC BY-NC Atıf-GayriTicari lisansı ile açık erişime sunulur. Tüm yazarlar ve diğer dergi kullanıcıları bu durumu kabul etmiş sayılırlar. CC BY-NC lisansı hakkında detaylı bilgiye erişmek için tıklayınız.